Calibration of thermocouples from 419,527 °C (freezing point of Zn) up to 1492 °C (melting point of the Pd-C eutectic), by the temperature fixed point and comparison methods

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F20%3AN0000017" target="_blank" >RIV/00177016:_____/20:N0000017 - isvavai.cz</a>

  • Result on the web

    <a href="https://iopscience.iop.org/article/10.1088/0026-1394/57/1A/03006" target="_blank" >https://iopscience.iop.org/article/10.1088/0026-1394/57/1A/03006</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1088/0026-1394/57/1A/03006" target="_blank" >10.1088/0026-1394/57/1A/03006</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Calibration of thermocouples from 419,527 °C (freezing point of Zn) up to 1492 °C (melting point of the Pd-C eutectic), by the temperature fixed point and comparison methods

  • Original language description

    This report present the results of the EURAMET inter-comparison carried out to compare the calibration of thermocouples from 419,527 °C (freezing point of Zn) up to 1492 °C (melting point of the Pd-C eutectic). This inter-comparison is intended to be used to support the calibration and measurement capabilities (CMCs) of the participants in the calibration of thermocouples. The comparison was organized in three loops with nineteen participating laboratories and it allowed the performance of the measurement either in fixed points and/or by comparison. The method used to analyse the results of the comparison was the generalized weighted mean that takes into account the full covariance matrix that includes correlations between the participants which have similar traceability sources and between measurements performed by the same laboratory (i.e. the pilots that performed measurements at the beginning and at the end of each loop and the measurements performed by the same laboratory in fixed points and by comparison at the same calibration point). Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database

  • CEP classification

  • OECD FORD branch

    20303 - Thermodynamics

Result continuities

  • Project

  • Continuities

    I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace

Others

  • Publication year

    2020

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Metrologia

  • ISSN

    0026-1394

  • e-ISSN

    1681-7575

  • Volume of the periodical

    57

  • Issue of the periodical within the volume

    1

  • Country of publishing house

    GB - UNITED KINGDOM

  • Number of pages

    121

  • Pages from-to

    1-121

  • UT code for WoS article

    000627255600024