X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F25%3AN0000109" target="_blank" >RIV/00177016:_____/25:N0000109 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.5114/jcb.2025.154104" target="_blank" >https://doi.org/10.5114/jcb.2025.154104</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.5114/jcb.2025.154104" target="_blank" >10.5114/jcb.2025.154104</a>
Alternative languages
Result language
angličtina
Original language name
X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer
Original language description
Purpose: The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50. Material and methods: Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis. Results: The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed. Conclusions: The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10301 - Atomic, molecular and chemical physics (physics of atoms and molecules including collision, interaction with radiation, magnetic resonances, Mössbauer effect)
Result continuities
Project
<a href="/en/project/8B19003" target="_blank" >8B19003: Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices</a><br>
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Contemporary Brachytherapy
ISSN
1689-832X
e-ISSN
2081-2841
Volume of the periodical
17
Issue of the periodical within the volume
4
Country of publishing house
PL - POLAND
Number of pages
8
Pages from-to
—
UT code for WoS article
001584231200004
EID of the result in the Scopus database
—