Patient skin dosimetry in interventional cardiology in the Czech Republic
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00179906%3A_____%2F11%3A10110626" target="_blank" >RIV/00179906:_____/11:10110626 - isvavai.cz</a>
Alternative codes found
RIV/86652052:_____/11:#0000037 RIV/68407700:21340/11:00228656
Result on the web
<a href="http://rpd.oxfordjournals.org/content/147/1-2/106.abstract" target="_blank" >http://rpd.oxfordjournals.org/content/147/1-2/106.abstract</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncr284" target="_blank" >10.1093/rpd/ncr284</a>
Alternative languages
Result language
angličtina
Original language name
Patient skin dosimetry in interventional cardiology in the Czech Republic
Original language description
In this study, skin dosimetry of patients undergoing interventional cardiology procedures is presented. Three hospitals were included. Two methods were used for skin dosimetry-radiochromic dosimetry films and reconstruction of skin dose distribution based on examination protocol. Maximum skin doses (MSD) obtained from both methods were compared for 175 patients. For patients for whom the film MSD was } 1 Gy, the reconstruction MSD differed from the film MSD in the range of +/- 50 % for 83 % of patients.For remaining patients, the difference was higher and it was caused by longer fluoroscopy time. For 59 patients for whom the cumulative dose was known, the cumulative dose was compared with the film MSD. Skin dosimetry with radiochromic films is more accurate than the reconstruction method, but films do not include X-ray fields from lateral projections whilest reconstructions do.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FA - Cardiovascular diseases including cardio-surgery
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/JC_4%2F2008" target="_blank" >JC_4/2008: Radiation Burden of Patients in Interventional Radiology.</a><br>
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2011
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Volume of the periodical
147
Issue of the periodical within the volume
1
Country of publishing house
GB - UNITED KINGDOM
Number of pages
5
Pages from-to
106-110
UT code for WoS article
000296355500025