Novel electrochemical detector for the determination of 8-hydroxy-2'-deoxyguanosine by HPLC
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11120%2F02%3A00011076" target="_blank" >RIV/00216208:11120/02:00011076 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Novel electrochemical detector for the determination of 8-hydroxy-2'-deoxyguanosine by HPLC
Original language description
A simple amperometric detector was tested for the HPLC determination of 8-hydroxy-2'-deoxyguanosine (8-OHdG) as the biomarker of in vivo oxidative DNA damage. The limit of detection was found to be 9.8 nmmol 8-OHdG/L sample injected. The level of 8-OHdGwas determined after SPE extraction of urine sample.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
CE - Biochemistry
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/GA203%2F01%2F0653" target="_blank" >GA203/01/0653: New electrochemical methods for quantitative detection of 8-hydroxy-2´-deoxyguanosine in urine</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2002
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
Proceedings of the 25th International Symposium on Capillary Chromatography
ISBN
—
ISSN
—
e-ISSN
—
Number of pages
1
Pages from-to
"O1"-"O6"
Publisher name
R.I.C.
Place of publication
Kortrijk
Event location
—
Event date
—
Type of event by nationality
—
UT code for WoS article
—