A framework for alignment of on-board imagers of medical linear accelerators
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F18%3A00102778" target="_blank" >RIV/00216224:14310/18:00102778 - isvavai.cz</a>
Result on the web
<a href="https://www.sciencedirect.com/science/article/pii/S1120179718300401" target="_blank" >https://www.sciencedirect.com/science/article/pii/S1120179718300401</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.ejmp.2018.02.017" target="_blank" >10.1016/j.ejmp.2018.02.017</a>
Alternative languages
Result language
angličtina
Original language name
A framework for alignment of on-board imagers of medical linear accelerators
Original language description
Purpose: To develop an assumption-free methodology for aligning the geometry of on-board imagers with the geometry of medical linear accelerators applied in image-guided radiotherapy (IGRT). Material: Alignment of the on-board imaging (OBI) system with respect to the accelerator system is achieved using a multi-modular phantom described elsewhere (Tabor et al., 2017), enabling the geometry of the linear accelerator to be specified without any pre-assumptions. Methods: The placement of two isocentres (of the on-board imager and of the therapeutic system) and of three frames of reference (those of the on-board imager, of the therapeutic system, and of the treatment table) is formulated as an optimization problem. It is solved by analysing the images of fiducial points positioned in 3D space in phantom modules attached to the treatment table and to the collimator of the accelerator. Fiducials are projected onto an imaging plane of unknown characteristics from a virtual source of unknown coordinates. Conclusions: An analytical framework exploiting projection images of the proposed multi-modular phantom has been developed, enabling precise alignment of the reference frame related to the on-board imager with the reference frame related to the therapeutic system. Within the proposed framework, the necessary corrections of treatment table positioning prior to patient irradiation, are delivered in the treatment table coordinates.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30224 - Radiology, nuclear medicine and medical imaging
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2018
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Physica Medica : European Journal of Medical Physics
ISSN
1120-1797
e-ISSN
—
Volume of the periodical
47
Issue of the periodical within the volume
2018
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
6
Pages from-to
80-85
UT code for WoS article
000428926700011