Assessment of MR stereotactic imaging and image co-registration accuracy for 3 different MR scanners by 3 different methods/phantoms: phantom and patient study
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F18%3A00007821" target="_blank" >RIV/00023884:_____/18:00007821 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.3171/2018.3.JNS18110" target="_blank" >http://dx.doi.org/10.3171/2018.3.JNS18110</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3171/2018.3.JNS18110" target="_blank" >10.3171/2018.3.JNS18110</a>
Alternative languages
Result language
angličtina
Original language name
Assessment of MR stereotactic imaging and image co-registration accuracy for 3 different MR scanners by 3 different methods/phantoms: phantom and patient study
Original language description
OBJECTIVE The aim of this study was to compare 3 different methods to assess the geometrical distortion of two 1.5- T and one 3-T magnetic resonance (MR) scanners and to evaluate co-registration accuracy. The overall uncertainty of each particular method was also evaluated. METHODS Three different MR phantoms were used: 2 commercial CIRS skull phantoms and PTGR known target phantom and 1 custom cylindrical Perspex phantom made in-house. All phantoms were fixed in the Leksell stereotactic frame and examined by a Siemens Somatom CT unit, two 1.5-T Siemens (Avanto and Symphony) MRI systems, and one 3-T Siemens (Skyra) MRI system. The images were evaluated using Leksell GammaPlan software, and geometrical deviation of the selected points from the reference values were determined. The deviations were further investigated for both definitions including fiducial-based and co-registration–based in the case of the CIRS phantom images. The same co-registration accuracy assessment was also performed for a clinical case. Patient stereotactic imaging was done on 3-T Skyra, 1.5-T Avanto, and CT scanners. RESULTS The accuracy of the CT scanner was determined as 0.10, 0.30, and 0.30 mm for X, Y, and Z coordinates, respectively. The total estimated uncertainty in distortion measurement in one coordinate was determined to be 0.32 mm and 0.14 mm, respectively, for methods using and not using CT as reference imaging. Slightly more significant distortions were observed when using the 3-T than either 1.5-T MR units. However, all scanners were comparable within the estimated measurement error. Observed deviation/distortion for individual X, Y, and Z stereotactic coordinates was typically within 0.50 mm for all 3 scanners and all 3 measurement methods employed. The total radial deviation/distortion was typically within 1.00 mm. Maximum total radial distortion was observed when the CIRS phantom was used; 1.08 ± 0.49 mm, 1.15 ± 0.48 mm, and 1.35 ± 0.49 mm for Symphony, Avanto, and Skyra, respectively. The co-registration process improved image stereotactic definition in a clinical case in which fiducial-based stereotactic definition was not accurate; this was demonstrated for 3-T stereotactic imaging in this study. The best results were shown for 3-T MR image coregistration with CT images improving image stereotactic definition by about 0.50 mm. The results obtained with patient data provided a similar trend of improvement in stereotactic definition by co-registration. CONCLUSIONS All 3 methods/phantoms used were evaluated as satisfactory for the image distortion measurement. The method using the PTGR phantom had the lowest uncertainty as no reference CT imaging was needed. Image coregistration can improve stereotactic image definition when fiducial-based definition is not accurate.
Czech name
—
Czech description
—
Classification
Type
J<sub>SC</sub> - Article in a specialist periodical, which is included in the SCOPUS database
CEP classification
—
OECD FORD branch
30103 - Neurosciences (including psychophysiology)
Result continuities
Project
<a href="/en/project/GA16-13323S" target="_blank" >GA16-13323S: MIcro and MAcro Connectomics of the Subthalamic nucleus in humans: impact of neuromodulation and dopamine depletion</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2018
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Neurosurgery
ISSN
0022-3085
e-ISSN
—
Volume of the periodical
129
Issue of the periodical within the volume
Suppl. 1
Country of publishing house
US - UNITED STATES
Number of pages
8
Pages from-to
125-132
UT code for WoS article
—
EID of the result in the Scopus database
2-s2.0-85058754673