Monolithic Capillary Column with an Integrated Electrochemical Detector
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216275%3A25310%2F17%3A39911330" target="_blank" >RIV/00216275:25310/17:39911330 - isvavai.cz</a>
Alternative codes found
RIV/00216224:14310/17:00113892
Result on the web
<a href="http://www.sciencedirect.com/science/article/pii/S0021967317309251" target="_blank" >http://www.sciencedirect.com/science/article/pii/S0021967317309251</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.chroma.2017.06.057" target="_blank" >10.1016/j.chroma.2017.06.057</a>
Alternative languages
Result language
angličtina
Original language name
Monolithic Capillary Column with an Integrated Electrochemical Detector
Original language description
The carbon fiber and silver microwire were used as working and pseudoreference electrode, respectively, and inserted into the ending of capillary to prepare monolithic capillary column with an integrated electrochemical detector. Prepared capillary devices offered stable and robust results with relative standard deviations of retention, resolution, and detection signal lower than 1.5, 5.5, and 5.0%, respectively. To further increase sensitivity of developed electrochemical microdetector, multiple pulse amperometry detection mode has been used. Optimized integrated device provided reliable chromatographic separation of mixture of neurotransmitters with calibration curve for dopamine linear from 0.5 to 20.0 mg L-1 and an instrumental limit of detection as low as 24 pg of injected dopamine. Finally, developed capillary column was applied to successful determination of dopamine in a human urine. By using both calibration curve and standard addition method, the dopamine level was determined to be 0.74 +/- 0.03 mgL(-1) and 0.71 +/- 0.02 mgL(-1), respectively. Triplicates of dopamine analysis provided relative standard deviations lower than 2.7% for intraday analyses, while interday relative standard deviations were lower than 3.6% for five consecutive days.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10406 - Analytical chemistry
Result continuities
Project
<a href="/en/project/GA14-22426S" target="_blank" >GA14-22426S: Development of an multifunctional monolithic capillary column with integrated sample focusing, separation, and on-column electrochemical detection</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2017
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Volume of the periodical
1509
Issue of the periodical within the volume
August
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
5
Pages from-to
171-175
UT code for WoS article
000407868800020
EID of the result in the Scopus database
—