Monitoring of CRT-D devices during radiation therapy in vitro
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00843989%3A_____%2F16%3AE0105389" target="_blank" >RIV/00843989:_____/16:E0105389 - isvavai.cz</a>
Alternative codes found
RIV/47813059:19510/16:N0000119 RIV/61989100:27240/16:86099589
Result on the web
<a href="http://dx.doi.org/10.1186/s12938-016-0144-7" target="_blank" >http://dx.doi.org/10.1186/s12938-016-0144-7</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1186/s12938-016-0144-7" target="_blank" >10.1186/s12938-016-0144-7</a>
Alternative languages
Result language
angličtina
Original language name
Monitoring of CRT-D devices during radiation therapy in vitro
Original language description
Background: Using of active cardiac medical devices increases steadily. In Europe, there were 183 implants of ICD and 944 implants of PM, 119 of biventricular ICD and 41 of biventricular PM, all per million inhabitants in 2014. Healthcare environments, including radiotherapy treatment rooms, are considered challenging for these implantable devices. Exposure to radiation may cause the device to experience premature elective replacement indicator, decreased pacing amplitude or pacing inhibition, inappropriate shocks or inhibition of tachyarrhythmia therapy and loss of device function. These impacts may be temporary or permanent. The aim of this study was to evaluate the influence of linear accelerator ionizing radiation dose of 10 Gy on the activity of the biventricular cardioverter-defibrillator in different position in radiation beam. Methods: Two identical wireless communication devices with all three leads were used for the measurement. Both systems were soused into solution saline and exposed in different position in the beam of linear accelerator per 10 Gy fractions. In comparison of usually used maximum recommended dose of 2 Gy, the radiation doses used in test were five times higher. Using the simultaneous monitoring wireless communication between device and its programmer allowed watching of the devices activities, noise occurrence or drop of biventricular pacing on the programmer screen, observed by local television loop camera. Results: At any device position in radiation beam, there were no influences of the device activity at dose of 10 Gy neither a significant increase of a solution saline temperature in any of the measured positions of CRT-D systems in linear accelerator. Conclusions: The results of the study indicated, that the recommendation dose for treating the patients with implantable devices are too conservative and the risk of device failure is not so high. The systems can easily withstand the dose fractions of tens Gy, which would allow cur...
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FD - Oncology and haematology
OECD FORD branch
—
Result continuities
Project
—
Continuities
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Others
Publication year
2016
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
BioMedical Engineering OnLine
ISSN
1475-925X
e-ISSN
—
Volume of the periodical
15
Issue of the periodical within the volume
n. 29
Country of publishing house
GB - UNITED KINGDOM
Number of pages
8
Pages from-to
"p. 1-8"
UT code for WoS article
000371664100001
EID of the result in the Scopus database
—