Recent developments and applications of selected ion flow tube mass spectrometry (SIFT-MS)
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F25%3A00569152" target="_blank" >RIV/61388955:_____/25:00569152 - isvavai.cz</a>
Result on the web
<a href="https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/10.1002/mas.21835" target="_blank" >https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/10.1002/mas.21835</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/mas.21835" target="_blank" >10.1002/mas.21835</a>
Alternative languages
Result language
angličtina
Original language name
Recent developments and applications of selected ion flow tube mass spectrometry (SIFT-MS)
Original language description
Selected ion flow tube mass spectrometry (SIFT-MS) is now recognized as the most versatile analytical technique for the identification and quantification of trace gases down to the parts-per-trillion by volume, pptv, range. This statement is supported by the wide reach of its applications, from real-time analysis, obviating sample collection of very humid exhaled breath, to its adoption in industrial scenarios for air quality monitoring. This review touches on the recent extensions to the underpinning ion chemistry kinetics library and the alternative challenge of using nitrogen carrier gas instead of helium. The addition of reagent anions in the Voice200 series of SIFT-MS instruments has enhanced the analytical capability, thus allowing analyses of volatile trace compounds in humid air that cannot be analyzed using reagent cations alone, as clarified by outlining the anion chemistry involved. Case studies are reviewed of breath analysis and bacterial culture volatile organic compound (VOC), emissions, environmental applications such as air, water, and soil analysis, workplace safety such as transport container fumigants, airborne contamination in semiconductor fabrication, food flavor and spoilage, drugs contamination and VOC emissions from packaging to demonstrate the stated qualities and uniqueness of the new generation SIFT-MS instrumentation. Finally, some advancements that can be made to improve the analytical capability and reach of SIFT-MS are mentioned.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10403 - Physical chemistry
Result continuities
Project
<a href="/en/project/GA21-25486S" target="_blank" >GA21-25486S: Selected ion flow drift tube mass spectrometry with negative ions and nitrogen carrier gas</a><br>
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Mass Spectrometry Reviews
ISSN
0277-7037
e-ISSN
1098-2787
Volume of the periodical
44
Issue of the periodical within the volume
2
Country of publishing house
US - UNITED STATES
Number of pages
34
Pages from-to
101-134
UT code for WoS article
000929499300001
EID of the result in the Scopus database
2-s2.0-85148043441