Noninvasive fetal heart rate monitoring: Validation of phonocardiography-based fiber-optic sensing and adaptive filtering using the nlms algorithm
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F17%3A10236101" target="_blank" >RIV/61989100:27240/17:10236101 - isvavai.cz</a>
Alternative codes found
RIV/00216224:14110/17:00100124 RIV/65269705:_____/17:00068499
Result on the web
<a href="https://www.scopus.com/record/display.uri?eid=2-s2.0-85030549140&origin=resultslist&sort=plf-f&src=s&sid=181567d2da1a33f250897a6fefa5be8d&sot=autdocs&sdt=autdocs&sl=18&s=AU-ID%2857014879400%29&relpos=4&citeCnt=0&searchTerm=" target="_blank" >https://www.scopus.com/record/display.uri?eid=2-s2.0-85030549140&origin=resultslist&sort=plf-f&src=s&sid=181567d2da1a33f250897a6fefa5be8d&sot=autdocs&sdt=autdocs&sl=18&s=AU-ID%2857014879400%29&relpos=4&citeCnt=0&searchTerm=</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.15598/aeee.v15i3.2195" target="_blank" >10.15598/aeee.v15i3.2195</a>
Alternative languages
Result language
angličtina
Original language name
Noninvasive fetal heart rate monitoring: Validation of phonocardiography-based fiber-optic sensing and adaptive filtering using the nlms algorithm
Original language description
Here we present the evaluation results of our novel noninvasive phonocardiographic-based fiber-optic sensor for fetal Heart Rate (fHR) detection using adaptive filtering and the NLMS Algorithm. The sensor uses two interferometric probes encapsulated inside a PolyDiMethylSiloxane (PDMS) polymer. Based on real data acquired from pregnant women in a suitable research laboratory environment, once they had given their written informed consents, we created a simplified dynamic signal model of the distribution of maternal and fetal heart sounds inside the maternal body. Building upon this signal model, we verified the functionality of our novel fiber-optic sensor and its associated adaptive filtering system using the NLMS Algorithm. The main reason why we chose this technology to develop our system was that it allows monitoring the fHR without exposing the fetus to any external energies or radiation (in contrast to the ultrasound-based Cardiotocography Method). We used objective criteria such as: Signal to Noise Ratios: SNRin, SNRout and Percentage Root-mean-square Difference (PRD) for our evaluations.
Czech name
—
Czech description
—
Classification
Type
J<sub>SC</sub> - Article in a specialist periodical, which is included in the SCOPUS database
CEP classification
—
OECD FORD branch
20201 - Electrical and electronic engineering
Result continuities
Project
<a href="/en/project/EE2.3.20.0217" target="_blank" >EE2.3.20.0217: The Development of Excellence of the Telecommunication Research Team in Relation to International Cooperation</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Others
Publication year
2017
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Advances in Electrical and Electronic Engineering
ISSN
1336-1376
e-ISSN
—
Volume of the periodical
15
Issue of the periodical within the volume
3
Country of publishing house
CZ - CZECH REPUBLIC
Number of pages
8
Pages from-to
544-552
UT code for WoS article
—