A Reversed-Phase High-Performance Liquid Chromatographic/Mass Spectrometric (RP-HPLC/MS-SIM) Method for Separation of 4-methylimidazole and 2-acetyl-4-(1,2,3,4-tetrahydroxybutyl)imidazole at pg levels
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62156489%3A43210%2F03%3A00000995" target="_blank" >RIV/62156489:43210/03:00000995 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
A Reversed-Phase High-Performance Liquid Chromatographic/Mass Spectrometric (RP-HPLC/MS-SIM) Method for Separation of 4-methylimidazole and 2-acetyl-4-(1,2,3,4-tetrahydroxybutyl)imidazole at pg levels
Original language description
A high-performance liquid chromatographic/mass spectrometric method (HPLC/MS) has been developed to determine both 4-methylimidazole (4MeI) and 2-acetyl-4-(1,2,3,4-tetrahydroxybutyl)imidazole (THI) in one run. Among three sorbents tested, the best peak shape of 4MeI was achieved on a reversed phase MetaChem Polaris C18-A at pH 9.5. The sensitivity and the range of UV detection at 215 and 290 nm for 4MeI and THI, respectively, were compared to the parameters achieved by the electrospray ionization mass spectrometric (ESI/MS) detection in the presence of 5 mmol.L-1 ammonium hydroxide using selected ion-monitoring (SIM) mode. The on-column limits of detection were 0.147 ng of 4MeI and 0.084 ng of THI using UV detection and 1 pg of 4MeI and 3 pg of THI using ESI/MS detection, respectively. The effect of both ammonium ion concentration and energy of fragmentation on 4MeI and THI ionization is discussed. The method could be applied for a fast and sensitive determination of 4MeI and THI in di
Czech name
RP HPLC/MS stanovení THI a 4MeI
Czech description
A high-performance liquid chromatographic/mass spectrometric method (HPLC/MS) has been developed to determine both 4-methylimidazole (4MeI) and 2-acetyl-4-(1,2,3,4-tetrahydroxybutyl)imidazole (THI) in one run. Among three sorbents tested, the best peak shape of 4MeI was achieved on a reversed phase MetaChem Polaris C18-A at pH 9.5. The sensitivity and the range of UV detection at 215 and 290 nm for 4MeI and THI, respectively, were compared to the parameters achieved by the electrospray ionization mass spectrometric (ESI/MS) detection in the presence of 5 mmol.L-1 ammonium hydroxide using selected ion-monitoring (SIM) mode. The on-column limits of detection were 0.147 ng of 4MeI and 0.084 ng of THI using UV detection and 1 pg of 4MeI and 3 pg of THI using ESI/MS detection, respectively. The effect of both ammonium ion concentration and energy of fragmentation on 4MeI and THI ionization is discussed. The method could be applied for a fast and sensitive determination of 4MeI and THI in di
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
CB - Analytical chemistry, separation
OECD FORD branch
—
Result continuities
Project
—
Continuities
Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2003
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Volume of the periodical
477
Issue of the periodical within the volume
1
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
10
Pages from-to
49-58
UT code for WoS article
—
EID of the result in the Scopus database
—