Experience with calibration of ionisation chambers which are used for absorbed dose determination in radiotherapeutical x-ray beams in Czech Republic.
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F63108089%3A_____%2F01%3A00007925" target="_blank" >RIV/63108089:_____/01:00007925 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Experience with calibration of ionisation chambers which are used for absorbed dose determination in radiotherapeutical x-ray beams in Czech Republic.
Original language description
Experience with calibration of ionisation chambers which are used for absorbed dose determination in radiotherapy X-ray beams in Czech Republic is presented. Calibration was performed in terms of air kerma at qualities from 5 keV (30 kV) to 160 keV (300kV) at Isovolt 400. Within a given chamber type, chamber-to-chamber variations in energy response can be significant. Long term energy dependence stability of the tested ionisation chambers is sufficient for the absorbed dose determination in the therapeutic X-ray beam.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
FD - Oncology and haematology
OECD FORD branch
—
Result continuities
Project
—
Continuities
Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2001
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
sborník - XXIV. Days of Radiation Protection
ISBN
80-88806-26-27
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
74-77
Publisher name
Žilinská Univerzita
Place of publication
Žilina
Event location
Jasná pod Chopkom, Slovensko
Event date
Nov 26, 2001
Type of event by nationality
EUR - Evropská akce