Two-parametric prescan calibration of gradient-induced sampling errors for rosette MRI
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F25%3A00636803" target="_blank" >RIV/68081731:_____/25:00636803 - isvavai.cz</a>
Result on the web
<a href="https://onlinelibrary.wiley.com/doi/10.1002/mrm.30355" target="_blank" >https://onlinelibrary.wiley.com/doi/10.1002/mrm.30355</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/mrm.30355" target="_blank" >10.1002/mrm.30355</a>
Alternative languages
Result language
angličtina
Original language name
Two-parametric prescan calibration of gradient-induced sampling errors for rosette MRI
Original language description
Purpose: The aim of this study was to develop a simple, robust, and easy-to-use calibration procedure for correcting misalignments in rosette MRI k-space sampling, with the objective of producing images with minimal artifacts. Methods: Quick automatic calibration scans were proposed for the beginning of the measurement to collect information on the time course of the rosette acquisition trajectory. A two-parameter model was devised to match the measured time-varying readout gradient delays and approximate the actual rosette sampling trajectory. The proposed calibration approach was implemented, and performance assessment was conducted on both phantoms and human subjects. Results: The fidelity of phantom and in vivo images exhibited significant improvement compared with uncorrected rosette data. The two-parameter calibration approach also demonstrated enhanced precision and reliability, as evidenced by quantitative relaxometry analyses. Conclusion: Adequate correction of data sampling is a crucial step in rosette MRI. The presented experimental results underscore the robustness, ease of implementation, and suitability for routine experimental use of the proposed two-parameter rosette trajectory calibration approach.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
20601 - Medical engineering
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Magnetic Resonance in Medicine
ISSN
0740-3194
e-ISSN
1522-2594
Volume of the periodical
93
Issue of the periodical within the volume
3
Country of publishing house
US - UNITED STATES
Number of pages
13
Pages from-to
1285-1297
UT code for WoS article
001339752700001
EID of the result in the Scopus database
2-s2.0-85207302468