Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma.
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F01%3A02020354" target="_blank" >RIV/68378271:_____/01:02020354 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma.
Original language description
Responses of thermoluminescet dosimeters of CaF 2 :Dy, CaF 2 :Mn, and LiF:Mg,Cu,P to the deposited energy of monochromatic 1.75-3.5 keV radiation were measured with respect to an increase in the local dose due to the attenuation length that was much shorter than the TLD thickness. The responses of TLDs to X-ray emission from the laser-produced aluminium plasma were found to be independent of the dose rate within a wide range.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
BL - Plasma physics and discharge through gases
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/LA%20055" target="_blank" >LA 055: Research within the Framework of Internationale Centre of Dense Magnetized Plasma</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2001
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Nukleonika
ISSN
0029-5922
e-ISSN
—
Volume of the periodical
46
Issue of the periodical within the volume
Suppl.
Country of publishing house
PL - POLAND
Number of pages
3
Pages from-to
"S49"-"S51"
UT code for WoS article
—
EID of the result in the Scopus database
—