A novel tool for the absolute end-to-end calibration of fluorescence telescopes - The XY-Scanner
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F24%3A00636111" target="_blank" >RIV/68378271:_____/24:00636111 - isvavai.cz</a>
Result on the web
<a href="https://pos.sissa.it/444/305/pdf" target="_blank" >https://pos.sissa.it/444/305/pdf</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.22323/1.444.0305" target="_blank" >10.22323/1.444.0305</a>
Alternative languages
Result language
angličtina
Original language name
A novel tool for the absolute end-to-end calibration of fluorescence telescopes - The XY-Scanner
Original language description
The Pierre Auger Observatory uses 27 large-aperture wide-angle Schmidt telescopes to measure the longitudinal profile of air showers using the air-fluorescence technique. Up to the year 2013, the absolute calibration of the telescopes was performed by mounting a uniform large-diameter light source on each of the telescopes and illuminating the entire aperture with a known photon flux. Due to the high amount of work and person-power required, this procedure was only carried out roughly once every three years, and a relative calibration was performed every night to track short-term changes. Since 2013, only the relative calibration has been performed. In this paper, we present a novel tool for the absolute end-to-end calibration of the fluorescence detectors, the XY-Scanner. The XY-Scanner uses a portable integrating sphere as a light source, which has been absolutely calibrated. This light source is installed onto a motorized rail system and moved across the aperture of each telescope. We mimic the illumination of the entire aperture by flashing the light source at ∼1700 positions evenly distributed across the telescope aperture. For the absolute calibration of the light source, we built a dedicated setup that uses a NIST-calibrated photodiode to measure the average photon flux and a PMT to track the pulse-to-pulse stability. We present the laboratory setups used to study the characteristics of the employed light sources and discuss the inter-calibration between selected telescopes.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
—
OECD FORD branch
10303 - Particles and field physics
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2024
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
Proceedings of Science
ISBN
—
ISSN
1824-8039
e-ISSN
—
Number of pages
13
Pages from-to
305
Publisher name
Sissa Medilab srl
Place of publication
Trieste
Event location
Nagoya
Event date
Jul 26, 2023
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—