Numerical Analyses of the Gyroid Structure for Use in Dental and Orthopaedic Implants
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21110%2F19%3A00334145" target="_blank" >RIV/68407700:21110/19:00334145 - isvavai.cz</a>
Alternative codes found
RIV/68407700:21110/19:00342950
Result on the web
<a href="http://experimentalni-mechanika.cz/cs/konference/konference/2019.html" target="_blank" >http://experimentalni-mechanika.cz/cs/konference/konference/2019.html</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Numerical Analyses of the Gyroid Structure for Use in Dental and Orthopaedic Implants
Original language description
The proposed work deals with the numerical analyses of different variants of the gyroid structure. The analyses were carried out to determine the optimal geometry (wall thickness, spacing) of the gyroid structure for use in dental and orthopaedic implants. Since the human body represents an environment of (compared to metals) soft tissues, it is not always desirable to introduce a structure with high stiffness as an implant. The gyroid structure (on the macroscopical level) represents a porous structure that reduces the overall stiffness of the implant if applied on the outer layer of the implant. What is more, this kind of mesh provides interesting features for tissue scaffolding due to high porosity which gives an opportunity to design a structure that can also meet the requirement for the optimal pore size with regard to the most effective bone ingrowth. Furthermore, we can minimize the stress shield effect which can lead to the bone resorption after implant insertion.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
—
OECD FORD branch
30404 - Biomaterials (as related to medical implants, devices, sensors)
Result continuities
Project
<a href="/en/project/TJ01000328" target="_blank" >TJ01000328: Optimization of 3D-printed trabecular structures for use in implant dentistry and their mechanical analyses</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
EAN 2019 57th conference on experimental stress analysis CONFERENCE PROCEEDINGS
ISBN
9788021457669
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
544-547
Publisher name
Česká společnost pro mechaniku
Place of publication
Praha
Event location
Luhačovice
Event date
Jun 3, 2019
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—