Determination of Absorbed Dose in the Experimental Animal Irradiated on the Leksell Gamma Knife
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21220%2F01%3A04068249" target="_blank" >RIV/68407700:21220/01:04068249 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Determination of Absorbed Dose in the Experimental Animal Irradiated on the Leksell Gamma Knife
Original language description
The purpose of this study was to evaluate geometric and dosimetric inaccuracies during the irradiation of the experimental animal (rat) brain on the Lekssel gamma knife. Altogetrher three types of dosimeters were employed for this measurements: a) thermoluminescent dosimeter, b) semiconductor detector, c) polymer-gel dosimeter BANG-2 type. Special glass phantom mimicking exactly the shape of the rat body and filled by the polymer gel was used for measurements of relative dose distribution and evaluationof geometric inaccuracies in the rat brain.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
BG - Nuclear, atomic and molecular physics, accelerators
OECD FORD branch
—
Result continuities
Project
—
Continuities
Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2001
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
24th Days of Radiation Protection
ISBN
80-88806-26-7
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
121-124
Publisher name
EDIS
Place of publication
Žilina
Event location
Jasná pod Chopkom
Event date
Nov 26, 2001
Type of event by nationality
WRD - Celosvětová akce