A stereotactic treatment procedure performed with Leksell gamma knife was verified. There was prepared a special head phantom filled with "home made" polymer-gel dosimeter and consequently underwent comlex stereotactic treatment procedure - MR localization, treatment planning and irradiation. MR slices obtained after dosimeter irradiation were used for 3D dose distribution map construction. The corresponding MR slices originated from treatment planning systém and from evaluated dosimeter were taken together in order to compare (dis)agreement of planned and measured isodoses. There was observed a good agreement of 3D dose distribution planned and measured.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
BG - Nuclear, atomic and molecular physics, accelerators
OECD FORD branch
—
Result continuities
Project
—
Continuities
Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2001
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
9th Scientific Meeting and Exhibition
ISBN
—
ISSN
—
e-ISSN
—
Number of pages
3
Pages from-to
428-430
Publisher name
University of Glasgow
Place of publication
Glasgow
Event location
Glasgow
Event date
Apr 21, 2001
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—