Measurement of Temperature Increase of Metal Hip Replacements During Magnetic Resonance Imaging
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F14%3A00220623" target="_blank" >RIV/68407700:21230/14:00220623 - isvavai.cz</a>
Alternative codes found
RIV/68407700:21460/14:00220623
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Measurement of Temperature Increase of Metal Hip Replacements During Magnetic Resonance Imaging
Original language description
In this paper measurements of temperature increase of three dierent hip replace- ments during MRI were carried out. The hip replacements were embedded in an agar phantom which mimics dielectric properties of muscle tissue and were placed into a MRI system with the static magnetic eld intensity of B0 = 1:5 T. The MRI of the samples took in total 20 minutes. Thermograms of the hip replacements were taken with a thermographic camera in two dierent time instances.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
JA - Electronics and optoelectronics
OECD FORD branch
—
Result continuities
Project
—
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2014
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
PIERS 2014 Guangzhou
ISBN
978-1-934142-28-8
ISSN
1559-9450
e-ISSN
—
Number of pages
4
Pages from-to
2746-2749
Publisher name
Electromagnetics Academy
Place of publication
Cambridge
Event location
Guangzhou
Event date
Aug 25, 2014
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—