All

What are you looking for?

All
Projects
Results
Organizations

Quick search

  • Projects supported by TA ČR
  • Excellent projects
  • Projects with the highest public support
  • Current projects

Smart search

  • That is how I find a specific +word
  • That is how I leave the -word out of the results
  • “That is how I can find the whole phrase”

Tidal Volume Dependence on the Ventilatory Frequency and Alveolar Compliance in HFOV

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F13%3A00210165" target="_blank" >RIV/68407700:21460/13:00210165 - isvavai.cz</a>

  • Result on the web

    <a href="http://www.degruyter.com/view/j/bmte.2013.58.issue-s1-N/bmt-2013-4329/bmt-2013-4329.xml?rskey=geRT6Y&result=3" target="_blank" >http://www.degruyter.com/view/j/bmte.2013.58.issue-s1-N/bmt-2013-4329/bmt-2013-4329.xml?rskey=geRT6Y&result=3</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1515/bmt-2013-4329" target="_blank" >10.1515/bmt-2013-4329</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Tidal Volume Dependence on the Ventilatory Frequency and Alveolar Compliance in HFOV

  • Original language description

    High frequency oscillatory ventilation has limited possibility of monitoring of intrapulmonary parameters. We have measured tidal volume during high frequency oscillatory ventilation by external monitor. The experimental circuit consisted of ventilator 3100 with its patient circuit, lung simulator 5600i and monitor Florian for measuring of tidal volume. Tidal volume was measured for different values of alveolar compliance that were set on the lung simulator. Measurement was repeated for ventilatory frequencies 3, 5, 7 and 10 Hz. Ventilatory parameters were held constant during the experiment. We have found that change of alveolar compliance has minimal effect on the delivered tidal volume.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    FS - Medical facilities, apparatus and equipment

  • OECD FORD branch

Result continuities

  • Project

  • Continuities

    S - Specificky vyzkum na vysokych skolach

Others

  • Publication year

    2013

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Biomedizinische Technik / Biomedical Engineering

  • ISSN

    0013-5585

  • e-ISSN

  • Volume of the periodical

    58

  • Issue of the periodical within the volume

    S1

  • Country of publishing house

    US - UNITED STATES

  • Number of pages

    2

  • Pages from-to

  • UT code for WoS article

    000329463600391

  • EID of the result in the Scopus database