Correction of measured X-ray spectra by an analytical response matrix using Matlab
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F15%3A%230000378" target="_blank" >RIV/86652052:_____/15:#0000378 - isvavai.cz</a>
Result on the web
<a href="http://www2.humusoft.cz/www/papers/tcp2015/037_kurkova.pdf" target="_blank" >http://www2.humusoft.cz/www/papers/tcp2015/037_kurkova.pdf</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Correction of measured X-ray spectra by an analytical response matrix using Matlab
Original language description
The response matrix R(Ej, Ek) of the Amptek CdTe XR-100T detector was constructed to enable numerical correction of the detection efficiency, escape effects and Compton scattering in the measured X-ray spectra with energies up to 160 keV. The incomplete charge collection was not considered. The matrix R was constructed analytically and can be used for spectra correction by a simple division of the vector representing the measured spectrum by this matrix. Presented at Technical Computing Prague 2015.
Czech name
—
Czech description
—
Classification
Type
O - Miscellaneous
CEP classification
FS - Medical facilities, apparatus and equipment
OECD FORD branch
—
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2015
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů