Effectiveness of single- vs dual-coil implantable defibrillator leads: An observational analysis from the SIMPLE study
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F19%3A00078262" target="_blank" >RIV/00023001:_____/19:00078262 - isvavai.cz</a>
Result on the web
<a href="https://onlinelibrary.wiley.com/doi/pdf/10.1111/jce.13943" target="_blank" >https://onlinelibrary.wiley.com/doi/pdf/10.1111/jce.13943</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1111/jce.13943" target="_blank" >10.1111/jce.13943</a>
Alternative languages
Result language
angličtina
Original language name
Effectiveness of single- vs dual-coil implantable defibrillator leads: An observational analysis from the SIMPLE study
Original language description
Introduction Dual-coil leads (DC-leads) were the standard of choice since the first nonthoracotomy implantable cardioverter/defibrillator (ICD). We used contemporary data to determine if DC-leads offer any advantage over single-coil leads (SC-leads), in terms of defibrillation efficacy, safety, clinical outcome, and complication rates. Methods and Results In the Shockless IMPLant Evaluation study, 2500 patients received a first implanted ICD and were randomized to implantation with or without defibrillation testing. Two thousand and four hundred seventy-five patients received SC-coil or DC-coil leads (SC-leads in 1025/2475 patients; 41.4%). In patients who underwent defibrillation testing (n = 1204), patients with both lead types were equally likely to achieve an adequate defibrillation safety margin (88.8% vs 91.2%; P = 0.16). There was no overall effect of lead type on the primary study endpoint of "failed appropriate shock or arrhythmic death" (adjusted HR 1.18; 95% CI, 0.86-1.62; P = 0.300), and on all-cause mortality (SC-leads: 5.34%/year; DC-leads: 5.48%/year; adjusted HR 1.16; 95% CI, 0.94-1.43; P = 0.168). However, among patients without prior heart failure (HF), and SC-leads had a significantly higher risk of failed appropriate shock or arrhythmic death (adjusted HR 7.02; 95% CI, 2.41-20.5). There were no differences in complication rates. Conclusion In this nonrandomized evaluation, there was no overall difference in defibrillation efficacy, safety, outcome, and complication rates between SC-leads and DC-leads. However, DC-leads were associated with a reduction in the composite of failed appropriate shock or arrhythmic death in the subgroup of non-HF patients. Considering riskier future lead extraction with DC-leads, SC-leads appears to be preferable in the majority of patients.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30201 - Cardiac and Cardiovascular systems
Result continuities
Project
—
Continuities
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of cardiovascular electrophysiology
ISSN
1045-3873
e-ISSN
—
Volume of the periodical
30
Issue of the periodical within the volume
7
Country of publishing house
US - UNITED STATES
Number of pages
8
Pages from-to
1078-1085
UT code for WoS article
000476777200010
EID of the result in the Scopus database
2-s2.0-85064686777