Primary standard for liquid flow rates between 30 and 1500 nl/min based on volume expansion

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F15%3A%230001087" target="_blank" >RIV/00177016:_____/15:#0001087 - isvavai.cz</a>

  • Result on the web

    <a href="http://www.degruyter.com/view/j/bmte.2015.60.issue-4/bmt-2014-0132/bmt-2014-0132.xml" target="_blank" >http://www.degruyter.com/view/j/bmte.2015.60.issue-4/bmt-2014-0132/bmt-2014-0132.xml</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1515/bmt-2014-0132" target="_blank" >10.1515/bmt-2014-0132</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Primary standard for liquid flow rates between 30 and 1500 nl/min based on volume expansion

  • Original language description

    An increasing number of microfluidic systems operate at flow rates below 1 ?l/min. Applications include (implanted) micropumps for drug delivery, liquid chromatography, and microreactors. For the applications where the absolute accuracy is important, a proper calibration is required. However, with standard calibration facilities, flow rate calibrations below ~1 ?l/min are not feasible because of a too large calibration uncertainty. In the current research, a traceable flow rate using a certain temperature increase rate is proposed. When the fluid properties, starting mass, and temperature increase rate are known, this principle yields a direct link to SI units, which makes it a primary standard. In this article, it will be shown that this principle enables flow rate uncertainties in the order of 2-3% for flow rates from 30 to 1500 nl/min.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    BK - Liquid mechanics

  • OECD FORD branch

Result continuities

  • Project

    <a href="/en/project/7AX13001" target="_blank" >7AX13001: Metrology for drug delivery</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace

Others

  • Publication year

    2015

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Biomedizinische Technik

  • ISSN

    0013-5585

  • e-ISSN

  • Volume of the periodical

    60

  • Issue of the periodical within the volume

    4

  • Country of publishing house

    DE - GERMANY

  • Number of pages

    18

  • Pages from-to

    317-335

  • UT code for WoS article