Modelling SPECT auto-contouring acquisitions for 177Lu & 131I molecular radiotherapy using new developments in Geant4/GATE
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F22%3AN0000034" target="_blank" >RIV/00177016:_____/22:N0000034 - isvavai.cz</a>
Result on the web
<a href="https://www.sciencedirect.com/science/article/pii/S1120179722019408" target="_blank" >https://www.sciencedirect.com/science/article/pii/S1120179722019408</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.ejmp.2022.03.003." target="_blank" >10.1016/j.ejmp.2022.03.003.</a>
Alternative languages
Result language
angličtina
Original language name
Modelling SPECT auto-contouring acquisitions for 177Lu & 131I molecular radiotherapy using new developments in Geant4/GATE
Original language description
Purpose: Monte Carlo modelling of SPECT imaging in Molecular Radiotherapy can improve activity quantification. Until now, SPECT modelling with GATE only considered circular orbit (CO) acquisitions. This cannot reproduce auto-contour acquisitions, where the detector head moves close to the patient to improve image resolution. The aim of this work is to develop and validate an auto-contouring step-and-shoot acquisition mode for GATE SPECT modelling. Methods: 177Lu and 131I SPECT experimental acquisitions performed on a Siemens Symbia T2 and GE Discovery 670 gamma camera, respectively, were modelled. SPECT projections were obtained for a cylindrical Jaszczak phantom and a lung and spine phantom. Detector head parameters (radial positions and acquisition angles) were extracted from the experimental projections to model the non-circular orbit (NCO) detector motion. The gamma camera model was validated against the experimental projections obtained with the cylindrical Jaszczak (177Lu) and lung and spine phantom (131I). Then, 177Lu and 131I CO and NCO SPECT projections were simulated to validate the impact of explicit NCO modelling on simulated projections. Results: Experimental and simulated SPECT images were compared using the gamma index, and were in good agreement with gamma index passing rate (GIPR) and gammaavg of 96.27%, 0.242 (177Lu) and 92.89%, 0.36 (131I). Then, simulated 177Lu and 131I CO and NCO SPECT projections were compared. The GIPR, gammaavg between the two gamma camera motions was 99.85%, 0.108 for 177Lu and 75.58%, 0.6 for 131I. Conclusion: This work thereby justifies the need for auto-contouring modelling for isotopes with high septal penetration. © 2022 Associazione Italiana di Fisica Medica e Sanitaria
Czech name
—
Czech description
—
Classification
Type
J<sub>SC</sub> - Article in a specialist periodical, which is included in the SCOPUS database
CEP classification
—
OECD FORD branch
30224 - Radiology, nuclear medicine and medical imaging
Result continuities
Project
<a href="/en/project/8B16002" target="_blank" >8B16002: Metrology for clinical implementation of dosimetry in molecular radiotherapy</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2022
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Physica Medica
ISSN
11201797
e-ISSN
—
Volume of the periodical
96
Issue of the periodical within the volume
April 2022
Country of publishing house
IT - ITALY
Number of pages
13
Pages from-to
101 -113
UT code for WoS article
—
EID of the result in the Scopus database