Electrocardiographic characterization of non-selective His-bundle pacing: validation of novel diagnostic criteria
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11120%2F19%3A43918895" target="_blank" >RIV/00216208:11120/19:43918895 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.1093/europace/euz275" target="_blank" >https://doi.org/10.1093/europace/euz275</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/europace/euz275" target="_blank" >10.1093/europace/euz275</a>
Alternative languages
Result language
angličtina
Original language name
Electrocardiographic characterization of non-selective His-bundle pacing: validation of novel diagnostic criteria
Original language description
AIMS: Permanent His-bundle (HB) pacing is usually accompanied by simultaneous capture of the adjacent right ventricular (RV) myocardium-this is described as a non-selective (ns)-HB pacing. It is of clinical importance to confirm HB capture using standard electrocardiogram (ECG). Our aim was to identify ECG criteria for loss of HB capture during ns-HB pacing. METHODS AND RESULTS: Patients with permanent HB pacing were recruited. Electrocardiograms during ns-HB pacing and loss of HB capture (RV-only capture) were obtained. Electrocardiogram criteria for loss/presence of HB capture were identified. In the validation phase, these criteria and the 'HB ECG algorithm' were tested using a separate, sizable set of ECGs. A total of 353 ECG (226 ns-HB and 128 RV-only) were obtained from 226 patients with permanent HB pacing devices. QRS notch/slur in left ventricular leads and R-wave peak time (RWPT) in lead V6 were identified as the best features for differentiation. The 'HB ECG algorithm' based on these features correctly classified 87.1% of cases with sensitivity and specificity of 93.2% and 83.9%, respectively. The criteria for definitive diagnosis of ns-HB capture (no QRS slur/notch in Leads I, V1, V4-V6, and the V6 RWPT LESS-THAN OR EQUAL TO 100 ms) presented 100% specificity. CONCLUSION: A novel ECG algorithm for the diagnosis of loss of HB capture and criteria for definitive confirmation of HB capture were formulated and validated. The algorithm might be useful during follow-up and the criteria for definitive confirmation of ns-HB capture offer a simple and reliable ancillary procedural endpoint during HB device implantation.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30201 - Cardiac and Cardiovascular systems
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Europace
ISSN
1099-5129
e-ISSN
—
Volume of the periodical
21
Issue of the periodical within the volume
12
Country of publishing house
GB - UNITED KINGDOM
Number of pages
8
Pages from-to
1857-1864
UT code for WoS article
000506802500020
EID of the result in the Scopus database
2-s2.0-85077995365