All

What are you looking for?

All
Projects
Results
Organizations

Quick search

  • Projects supported by TA ČR
  • Excellent projects
  • Projects with the highest public support
  • Current projects

Smart search

  • That is how I find a specific +word
  • That is how I leave the -word out of the results
  • “That is how I can find the whole phrase”

Short-and long-term electrical performance of the 5086MRI pacing lead

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15110%2F14%3A33148695" target="_blank" >RIV/61989592:15110/14:33148695 - isvavai.cz</a>

  • Result on the web

    <a href="http://dx.doi.org/10.1016/j.hrthm.2013.10.056" target="_blank" >http://dx.doi.org/10.1016/j.hrthm.2013.10.056</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1016/j.hrthm.2013.10.056" target="_blank" >10.1016/j.hrthm.2013.10.056</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Short-and long-term electrical performance of the 5086MRI pacing lead

  • Original language description

    Background:Owing to potential patient harm, magnetic resonance imaging (MRI) in patients with cardiac implantable electronic devices has traditionally been contraindicated. Electrical and handling characteristics of MR conditional pacing leads compared with those of traditional leads have not been established. Objective: To compare short- and long-term performance of the Medtronic 5086MRI lead with that of the Medtronic 5076 lead. Methods:Patients (n = 466) implanted with 2 5086MRI leads (EnRhythm MRI Study) and 316 patients implanted with 2 5076 leads in 2 prospective clinical investigations were analyzed. Results: Electrical characteristics were stable in both groups at implant and at 12 months and were clinically acceptable throughout. Ventricular capture thresholds were slightly higher at 12 months for the 5086MRI lead than for the 5076 lead (0.93 +- 0.47 V vs 0.74 +- 0.42 V; P { .001). Ventricular sensing amplitudes for the 5086MRI lead were lower at implant (9.0 +- 4.7 mV vs 13.9

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    FA - Cardiovascular diseases including cardio-surgery

  • OECD FORD branch

Result continuities

  • Project

  • Continuities

    N - Vyzkumna aktivita podporovana z neverejnych zdroju

Others

  • Publication year

    2014

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Heart Rhythm

  • ISSN

    1547-5271

  • e-ISSN

  • Volume of the periodical

    11

  • Issue of the periodical within the volume

    2

  • Country of publishing house

    US - UNITED STATES

  • Number of pages

    8

  • Pages from-to

    222-229

  • UT code for WoS article

  • EID of the result in the Scopus database