Organ and effective doses from verification techniques in Image-Guided Radiotherapy
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F11%3A00191466" target="_blank" >RIV/68407700:21340/11:00191466 - isvavai.cz</a>
Alternative codes found
RIV/86652052:_____/11:#0000032
Result on the web
<a href="http://rpd.oxfordjournals.org/cgi/reprint/ncr335?ijkey=ZCWsw7OMD7Z5ruS&keytype=ref" target="_blank" >http://rpd.oxfordjournals.org/cgi/reprint/ncr335?ijkey=ZCWsw7OMD7Z5ruS&keytype=ref</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncr335" target="_blank" >10.1093/rpd/ncr335</a>
Alternative languages
Result language
angličtina
Original language name
Organ and effective doses from verification techniques in Image-Guided Radiotherapy
Original language description
The purpose of this work was to evaluate organ doses and effective doses from three verification techniques in Image-Guided Radiotherapy: from kilovoltage (kV) cone beam CT (CBCT) scans, from two orthogonal kV images, and from two orthogonal megavoltage(MV) images for two different treatment sites: pelvis and head and neck (H&N). For comparison reasons, organ doses and effective doses from prostate and H&N radiotherapy were also evaluated. Measurements of organ doses were performed in a male anthropomorphic Rando phantom by means of thermoluminescent dosemeters (TLDs) MCP-N. In our investigation, measured organ doses from one CBCT scan, from two MV images, and from two kV images of pelvis represent typically 1 to 6 %, 1 to 10 %, and 0.05 to 1 %, respectively, of organ doses resulting from one fraction of prostate radiotherapy. The maximum effective doses from one CBCT scan, two kV images and two MV images of pelvis are 5.6 mSv, 0.8 mSv and 11.9 mSv, respectively.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FS - Medical facilities, apparatus and equipment
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/JC_4%2F2009" target="_blank" >JC_4/2009: Evaluation and verification of patient doses during advanced radiation therapy techniques (IMRT, IGRT).</a><br>
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2011
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Volume of the periodical
147
Issue of the periodical within the volume
1
Country of publishing house
GB - UNITED KINGDOM
Number of pages
4
Pages from-to
277-280
UT code for WoS article
000296355500065