Study of edgeless radiation detector with 3D spatial mapping technique

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F14%3A00229059" target="_blank" >RIV/68407700:21670/14:00229059 - isvavai.cz</a>

  • Result on the web

    <a href="http://iopscience.iop.org/1748-0221/9/04/C04004" target="_blank" >http://iopscience.iop.org/1748-0221/9/04/C04004</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1088/1748-0221/9/04/C04004" target="_blank" >10.1088/1748-0221/9/04/C04004</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Study of edgeless radiation detector with 3D spatial mapping technique

  • Original language description

    Edgeless radiation detector has gained increased attention due to its superiority in the defect-free edge fabrication and the capability to minimize the insensitive area at the detector edge. The doped edge in the edgeless detector is at the same potential with the back plane and causes a local distortion of the electric field at the detector edge. The deformed electric field alters the charge collection of the edge pixel and leads to an inaccurate charge interpolation. To study the influence of activeedges on the response of edge pixels, we used an advanced X-ray based 3D spatial mapping technique to visually show the charge collection volumes of pixels. Various edgeless detectors with diverse polarities, thicknesses and edge-to-pixel distances wereinvestigated. For the n-on-p (n+/p-/p+) edgeless detector, the mapping shows that the p-spray isolation method has the advantage of achieving a greater sensitive edge region compared to the p-stop method. And the p-on-p (p+/p-/n+) edgeles

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    BG - Nuclear, atomic and molecular physics, accelerators

  • OECD FORD branch

Result continuities

  • Project

    <a href="/en/project/TA01010237" target="_blank" >TA01010237: Facility for nondestructive testing, diagnostics and 3D imaging based on neutron radiography and tomography.</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2014

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Journal of Instrumentation

  • ISSN

    1748-0221

  • e-ISSN

  • Volume of the periodical

    9

  • Issue of the periodical within the volume

    9

  • Country of publishing house

    IT - ITALY

  • Number of pages

    9

  • Pages from-to

  • UT code for WoS article

    000336123800004

  • EID of the result in the Scopus database