Monte Carlo calibration of the whole-body counting detection system for in vivo measurement of people internally contaminated with 90Sr
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F86652052%3A_____%2F17%3AN0000018" target="_blank" >RIV/86652052:_____/17:N0000018 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1093/rpd/ncw324" target="_blank" >http://dx.doi.org/10.1093/rpd/ncw324</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncw324" target="_blank" >10.1093/rpd/ncw324</a>
Alternative languages
Result language
angličtina
Original language name
Monte Carlo calibration of the whole-body counting detection system for in vivo measurement of people internally contaminated with 90Sr
Original language description
This work is focused on numerical calibrations of the body counter for in vivo measurement of pure beta emitters through the produced bremsstrahlung radiation. Calibrations were performed using the UPh-02T block whole-body phantom and the Lawrence Livermore National Laboratory (LLNL) realistic torso phantom. Neither of these physical phantoms is appropriate for such calibrations; therefore, specific 90Sr sources have been manufactured to be used with the UPh-02T phantom for experimental measurement followed with Monte Carlo (MC) simulations. Calibrations with the LLNL torso phantom were carried out solely using MC technique. Different scenarios of the 90Sr distribution in the human tissues were considered for the spectrometer calibrations. MC simulations with the LLNL confirmed the applicability of the UPh-02T with specific 90Sr/90Y sources for experimental calibrations of the body counters for measurement of pure beta emitters. Differences in count rates in 50–200 keV for UPh-02T and LLNL were not >25% for all considered scenarios.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10511 - Environmental sciences (social aspects to be 5.7)
Result continuities
Project
<a href="/en/project/VF20102015014" target="_blank" >VF20102015014: Research of advanced methods of detection, estimation and management of resulting radioactive contamination</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2017
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Volume of the periodical
173
Issue of the periodical within the volume
1-3
Country of publishing house
GB - UNITED KINGDOM
Number of pages
7
Pages from-to
111-117
UT code for WoS article
000399768400021
EID of the result in the Scopus database
—