A novel radiofrequency ablation catheter using contact force sensing: Toccata study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F12%3A00055949" target="_blank" >RIV/00023001:_____/12:00055949 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.sciencedirect.com/science/article/pii/S1547527111009623" target="_blank" >http://www.sciencedirect.com/science/article/pii/S1547527111009623</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.hrthm.2011.08.021" target="_blank" >10.1016/j.hrthm.2011.08.021</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A novel radiofrequency ablation catheter using contact force sensing: Toccata study
Popis výsledku v původním jazyce
OBJECTIVES The aim of this multicenter study was to evaluate the device-and procedure-related safety of a novel force-sensing radiofrequency (RF) ablation catheter capable of measuring the real-time contact force (CF) and to present CF data and its possible implications on patient safety. BACKGROUND The clinical outcome of RF ablation for the treatment of cardiac arrhythmias may be affected by the CF between the catheter tip and the tissue. Insufficient CF may result in an ineffective lesion, whereas excessive CF may result in complications. METHODS Seventy-seven patients (43 with right-sided supraventricular tachycardia [SVT] and 34 with atrial fibrillation [AF]) received percutaneous ablation with the novel studied catheter. The CF applied and safetyevents related to the procedure were reported. RESULTS CF values at mapping ranged from 8 +/- 8 to 60 +/- 35 g and from 12 +/- 10 to 39 +/- 29 g in the SVT group and the LA group, respectively, showing a significant interinvestigator var
Název v anglickém jazyce
A novel radiofrequency ablation catheter using contact force sensing: Toccata study
Popis výsledku anglicky
OBJECTIVES The aim of this multicenter study was to evaluate the device-and procedure-related safety of a novel force-sensing radiofrequency (RF) ablation catheter capable of measuring the real-time contact force (CF) and to present CF data and its possible implications on patient safety. BACKGROUND The clinical outcome of RF ablation for the treatment of cardiac arrhythmias may be affected by the CF between the catheter tip and the tissue. Insufficient CF may result in an ineffective lesion, whereas excessive CF may result in complications. METHODS Seventy-seven patients (43 with right-sided supraventricular tachycardia [SVT] and 34 with atrial fibrillation [AF]) received percutaneous ablation with the novel studied catheter. The CF applied and safetyevents related to the procedure were reported. RESULTS CF values at mapping ranged from 8 +/- 8 to 60 +/- 35 g and from 12 +/- 10 to 39 +/- 29 g in the SVT group and the LA group, respectively, showing a significant interinvestigator var
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FA - Kardiovaskulární nemoci včetně kardiochirurgie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Heart Rhythm
ISSN
1547-5271
e-ISSN
—
Svazek periodika
9
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
6
Strana od-do
18-23
Kód UT WoS článku
000298640400005
EID výsledku v databázi Scopus
—