Multicentre clinical trial experience with the HeartMate 3 left ventricular assist device: 30-day outcomes
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F16%3A00060057" target="_blank" >RIV/00023001:_____/16:00060057 - isvavai.cz</a>
Výsledek na webu
<a href="http://ejcts.oxfordjournals.org/content/50/3/548" target="_blank" >http://ejcts.oxfordjournals.org/content/50/3/548</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/ejcts/ezw169" target="_blank" >10.1093/ejcts/ezw169</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Multicentre clinical trial experience with the HeartMate 3 left ventricular assist device: 30-day outcomes
Popis výsledku v původním jazyce
OBJECTIVES: The objective of this study was to describe the operative experience and 30-day outcomes of patients implanted with the HeartMate 3 Left Ventricular Assist System (LVAS) during the Conformite Europeenne (CE) Mark clinical trial. METHODS: Adult patients met inclusion and exclusion criteria defining advanced-stage heart failure and included the indications of bridge to transplant and destination therapy. Operative parameters, outcomes, adverse events, physical status and quality-of-life parameters were assessed in the first 30 days after LVAS implant. RESULTS: Fifty patients were implanted with the HeartMate 3 at 10 centres in 6 countries. The 30-day survival rate was 98%. The median operative and cardiopulmonary bypass times were 200 (range: 95-585) min and 84 (range: 47-250) min, respectively. Patients required transfusion with packed red blood cells (3.6 +/- 2.3 units), fresh frozen plasma (6.5 +/- 5 units) and platelets (2 +/- 1 units). Six patients (12%) required reoperation for postoperative bleeding and 10 patients (20%) did not require blood transfusion. The median intensive care time was 6 days (range: 1-112 days) and the total hospital stay was 28 days (range: 14-116 days). The most common adverse events were bleeding (15, 30%), arrhythmia (14, 28%) and infection (10, 20%). There were 2 (4%) strokes. CONCLUSIONS: The 30-day outcomes following implantation of the HeartMate 3 demonstrates excellent survival with low adverse event rates. The LVAD performed as intended with no haemolysis or device failure.
Název v anglickém jazyce
Multicentre clinical trial experience with the HeartMate 3 left ventricular assist device: 30-day outcomes
Popis výsledku anglicky
OBJECTIVES: The objective of this study was to describe the operative experience and 30-day outcomes of patients implanted with the HeartMate 3 Left Ventricular Assist System (LVAS) during the Conformite Europeenne (CE) Mark clinical trial. METHODS: Adult patients met inclusion and exclusion criteria defining advanced-stage heart failure and included the indications of bridge to transplant and destination therapy. Operative parameters, outcomes, adverse events, physical status and quality-of-life parameters were assessed in the first 30 days after LVAS implant. RESULTS: Fifty patients were implanted with the HeartMate 3 at 10 centres in 6 countries. The 30-day survival rate was 98%. The median operative and cardiopulmonary bypass times were 200 (range: 95-585) min and 84 (range: 47-250) min, respectively. Patients required transfusion with packed red blood cells (3.6 +/- 2.3 units), fresh frozen plasma (6.5 +/- 5 units) and platelets (2 +/- 1 units). Six patients (12%) required reoperation for postoperative bleeding and 10 patients (20%) did not require blood transfusion. The median intensive care time was 6 days (range: 1-112 days) and the total hospital stay was 28 days (range: 14-116 days). The most common adverse events were bleeding (15, 30%), arrhythmia (14, 28%) and infection (10, 20%). There were 2 (4%) strokes. CONCLUSIONS: The 30-day outcomes following implantation of the HeartMate 3 demonstrates excellent survival with low adverse event rates. The LVAD performed as intended with no haemolysis or device failure.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FA - Kardiovaskulární nemoci včetně kardiochirurgie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
European journal of cardio-thoracic surgery
ISSN
1010-7940
e-ISSN
—
Svazek periodika
50
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
548-554
Kód UT WoS článku
000386056400040
EID výsledku v databázi Scopus
—