Accuracy of electroanatomical mapping-guided cardiac radiotherapy for ventricular tachycardia: pitfalls and solutions
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F21%3A00081944" target="_blank" >RIV/00023001:_____/21:00081944 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00843989:_____/21:E0109287 RIV/00216208:11110/21:10435617
Výsledek na webu
<a href="https://academic.oup.com/europace/article/23/12/1989/6370658" target="_blank" >https://academic.oup.com/europace/article/23/12/1989/6370658</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/europace/euab195" target="_blank" >10.1093/europace/euab195</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Accuracy of electroanatomical mapping-guided cardiac radiotherapy for ventricular tachycardia: pitfalls and solutions
Popis výsledku v původním jazyce
Aims To analyse and optimize the interobserver agreement for gross target volume (GTV) delineation on cardiac computed tomography (CCT) based on electroanatomical mapping (EAM) data acquired to guide radiotherapy for ventricular tachycardia (VT). Methods and results Electroanatomical mapping data were exported and merged with the segmented CCT using manual registration by two observers. A GTV was created by both observers for predefined left ventricular (LV) areas based on preselected endocardial EAM points indicating a two-dimensional (2D) surface area of interest. The influence of (interobserver) registration accuracy and availability of EAM data on the final GTV and 2D surface location within each LV area was evaluated. The median distance between the CCT and EAM after registration was 2.7 mm, 95th percentile 6.2 mm for observer #1 and 3.0 mm, 95th percentile 7.6 mm for observer #2 (P = 0.9). Created GTVs were significantly different (8 vs. 19 mL) with lowest GTV overlap (35%) for lateral wall target areas. Similarly, the highest shift between 2D surfaces was observed for the septal LV (6.4 mm). The optimal surface registration accuracy (2.6 mm) and interobserver agreement (A interobserver EAM surface registration 1.3 mm) was achieved if at least three cardiac chambers were mapped, including high-quality endocardial LV EAM. Conclusion Detailed EAM of at least three chambers allows for accurate co-registration of EAM data with CCT and high interobserver agreement to guide radiotherapy of VT. However, the substrate location should be taken in consideration when creating a treatment volume margin. [GRAPHICS] .
Název v anglickém jazyce
Accuracy of electroanatomical mapping-guided cardiac radiotherapy for ventricular tachycardia: pitfalls and solutions
Popis výsledku anglicky
Aims To analyse and optimize the interobserver agreement for gross target volume (GTV) delineation on cardiac computed tomography (CCT) based on electroanatomical mapping (EAM) data acquired to guide radiotherapy for ventricular tachycardia (VT). Methods and results Electroanatomical mapping data were exported and merged with the segmented CCT using manual registration by two observers. A GTV was created by both observers for predefined left ventricular (LV) areas based on preselected endocardial EAM points indicating a two-dimensional (2D) surface area of interest. The influence of (interobserver) registration accuracy and availability of EAM data on the final GTV and 2D surface location within each LV area was evaluated. The median distance between the CCT and EAM after registration was 2.7 mm, 95th percentile 6.2 mm for observer #1 and 3.0 mm, 95th percentile 7.6 mm for observer #2 (P = 0.9). Created GTVs were significantly different (8 vs. 19 mL) with lowest GTV overlap (35%) for lateral wall target areas. Similarly, the highest shift between 2D surfaces was observed for the septal LV (6.4 mm). The optimal surface registration accuracy (2.6 mm) and interobserver agreement (A interobserver EAM surface registration 1.3 mm) was achieved if at least three cardiac chambers were mapped, including high-quality endocardial LV EAM. Conclusion Detailed EAM of at least three chambers allows for accurate co-registration of EAM data with CCT and high interobserver agreement to guide radiotherapy of VT. However, the substrate location should be taken in consideration when creating a treatment volume margin. [GRAPHICS] .
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
<a href="/cs/project/NU20-02-00244" target="_blank" >NU20-02-00244: Stereotaktická ablativní radiochirurgie rekurentní komorové tachykardie u strukturního onemocnění srdce</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Europace
ISSN
1099-5129
e-ISSN
—
Svazek periodika
23
Číslo periodika v rámci svazku
12
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
9
Strana od-do
1989-1997
Kód UT WoS článku
000733371300014
EID výsledku v databázi Scopus
2-s2.0-85122186404