Accuracy and feasibility of real-time continuous glucose monitoring in critically Ill patients after abdominal surgery and solid organ transplantation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F24%3A00084941" target="_blank" >RIV/00023001:_____/24:00084941 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216208:11110/24:10478288 RIV/00216208:11120/24:43926721
Výsledek na webu
<a href="https://diabetesjournals.org/care/article/47/6/956/154297/Accuracy-and-Feasibility-of-Real-time-Continuous" target="_blank" >https://diabetesjournals.org/care/article/47/6/956/154297/Accuracy-and-Feasibility-of-Real-time-Continuous</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.2337/dc23-1663" target="_blank" >10.2337/dc23-1663</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Accuracy and feasibility of real-time continuous glucose monitoring in critically Ill patients after abdominal surgery and solid organ transplantation
Popis výsledku v původním jazyce
OBJECTIVE Glycemia management in critical care is posing a challenge in frequent measuring and adequate insulin dose adjustment. In recent years, continuous glucose measurement has gained accuracy and reliability in outpatient and inpatient settings. The aim of this study was to assess the feasibility and accuracy of real-time continuous glucose monitoring (CGM) in ICU patients after major abdominal surgery. RESEARCH DESIGN AND METHODS We included patients undergoing pancreatic surgery and solid organ transplantation (liver, pancreas, islets of Langerhans, kidney) requiring an ICU stay after surgery. We used a Dexcom G6 sensor, placed in the infraclavicular region, for real-time CGM. Arterial blood glucose measured by the amperometric principle (ABL 800; Radiometer, Copenhagen, Denmark) served as a reference value and for calibration. Blood glucose was also routinely monitored by a StatStrip bedside glucose meter. Sensor accuracy was assessed by mean absolute relative difference (MARD), bias, modified Bland-Altman plot, and surveillance error grid for paired samples of glucose values from CGM and acid-base analyzer (ABL). RESULTS We analyzed data from 61 patients and obtained 1,546 paired glucose values from CGM and ABL. Active sensor use was 95.1%. MARD was 9.4%, relative bias was 1.4%, and 92.8% of values fell in zone A, 6.1% fell in zone B, and 1.2% fell in zone C of the surveillance error grid. Median time in range was 78%, with minimum (<1%) time spent in hypoglycemia. StatStrip glucose meter MARD compared with ABL was 5.8%. CONCLUSIONS Our study shows clinically applicable accuracy and reliability of Dexcom G6 CGM in postoperative ICU patients and a feasible alternative sensor placement site.
Název v anglickém jazyce
Accuracy and feasibility of real-time continuous glucose monitoring in critically Ill patients after abdominal surgery and solid organ transplantation
Popis výsledku anglicky
OBJECTIVE Glycemia management in critical care is posing a challenge in frequent measuring and adequate insulin dose adjustment. In recent years, continuous glucose measurement has gained accuracy and reliability in outpatient and inpatient settings. The aim of this study was to assess the feasibility and accuracy of real-time continuous glucose monitoring (CGM) in ICU patients after major abdominal surgery. RESEARCH DESIGN AND METHODS We included patients undergoing pancreatic surgery and solid organ transplantation (liver, pancreas, islets of Langerhans, kidney) requiring an ICU stay after surgery. We used a Dexcom G6 sensor, placed in the infraclavicular region, for real-time CGM. Arterial blood glucose measured by the amperometric principle (ABL 800; Radiometer, Copenhagen, Denmark) served as a reference value and for calibration. Blood glucose was also routinely monitored by a StatStrip bedside glucose meter. Sensor accuracy was assessed by mean absolute relative difference (MARD), bias, modified Bland-Altman plot, and surveillance error grid for paired samples of glucose values from CGM and acid-base analyzer (ABL). RESULTS We analyzed data from 61 patients and obtained 1,546 paired glucose values from CGM and ABL. Active sensor use was 95.1%. MARD was 9.4%, relative bias was 1.4%, and 92.8% of values fell in zone A, 6.1% fell in zone B, and 1.2% fell in zone C of the surveillance error grid. Median time in range was 78%, with minimum (<1%) time spent in hypoglycemia. StatStrip glucose meter MARD compared with ABL was 5.8%. CONCLUSIONS Our study shows clinically applicable accuracy and reliability of Dexcom G6 CGM in postoperative ICU patients and a feasible alternative sensor placement site.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30202 - Endocrinology and metabolism (including diabetes, hormones)
Návaznosti výsledku
Projekt
<a href="/cs/project/LX22NPO5104" target="_blank" >LX22NPO5104: Národní institut pro výzkum metabolických a kardiovaskulárních onemocnění</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Diabetes care
ISSN
0149-5992
e-ISSN
1935-5548
Svazek periodika
47
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
956-963
Kód UT WoS článku
001297160700015
EID výsledku v databázi Scopus
2-s2.0-85194013767