Temperature-Controlled Radiofrequency Ablation for Pulmonary Vein Isolation in Patients With Atrial Fibrillation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F12%3A00007550" target="_blank" >RIV/00023884:_____/12:00007550 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.jacc.2017.06.008" target="_blank" >http://dx.doi.org/10.1016/j.jacc.2017.06.008</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.jacc.2017.06.008" target="_blank" >10.1016/j.jacc.2017.06.008</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Temperature-Controlled Radiofrequency Ablation for Pulmonary Vein Isolation in Patients With Atrial Fibrillation
Popis výsledku v původním jazyce
Background Saline irrigation improved the safety of radiofrequency (RF) ablation, but the thermal feedback for energy titration is absent. Objectives To allow temperature-controlled irrigated ablation, a novel irrigated RF catheter was designed with a diamond-embedded tip (for rapid cooling) and 6 surface thermocouples to reflect tissue temperature. High-resolution electrograms (EGMs) from the split-tip electrode allowed rapid lesion assessment. The authors evaluated the preclinical and clinical performance of this catheter for pulmonary vein (PV) isolation. Methods Using the DiamondTemp (DT) catheter, pigs (n = 6) underwent discrete atrial ablation in a temperature control mode (60°C/50 W) until there was ∼80% EGM amplitude reduction. In a single-center clinical feasibility study, 35 patients underwent PV isolation with the DT catheter (study group); patients were planned for PV remapping after 3 months, regardless of symptomatology. A control group included 35 patients who underwent PV isolation with a standard force-sensing catheter. Results Porcine lesion histology revealed transmurality in 51 of 55 lesions (92.7%). In patients, all PVs were successfully isolated; no char or thrombus formation was observed. Compared with the control group, the study cohort had shorter mean RF application duration (26.3 ± 5.2 min vs. 89.2 ± 27.2 min; p < 0.001), shorter mean fluoroscopic time (11.2 ± 8.5 min vs. 19.5 ± 6.8 min; p < 0.001), and lower acute dormant PV reconduction (0 of 35 vs. 5 of 35; p = 0.024). At 3 months, 23 patients underwent remapping: 39 of 46 PV pairs (84.8%) remained durably isolated in 17 of these patients (73.9%). Conclusions This first-in-human series demonstrated that temperature-controlled irrigated ablation produced rapid, efficient, and durable PV isolation. (ACT DiamondTemp Temperature-Controlled and Contact Sensing RF Ablation Clinical Trial for Atrial Fibrillation [TRAC-AF]; NCT02821351)
Název v anglickém jazyce
Temperature-Controlled Radiofrequency Ablation for Pulmonary Vein Isolation in Patients With Atrial Fibrillation
Popis výsledku anglicky
Background Saline irrigation improved the safety of radiofrequency (RF) ablation, but the thermal feedback for energy titration is absent. Objectives To allow temperature-controlled irrigated ablation, a novel irrigated RF catheter was designed with a diamond-embedded tip (for rapid cooling) and 6 surface thermocouples to reflect tissue temperature. High-resolution electrograms (EGMs) from the split-tip electrode allowed rapid lesion assessment. The authors evaluated the preclinical and clinical performance of this catheter for pulmonary vein (PV) isolation. Methods Using the DiamondTemp (DT) catheter, pigs (n = 6) underwent discrete atrial ablation in a temperature control mode (60°C/50 W) until there was ∼80% EGM amplitude reduction. In a single-center clinical feasibility study, 35 patients underwent PV isolation with the DT catheter (study group); patients were planned for PV remapping after 3 months, regardless of symptomatology. A control group included 35 patients who underwent PV isolation with a standard force-sensing catheter. Results Porcine lesion histology revealed transmurality in 51 of 55 lesions (92.7%). In patients, all PVs were successfully isolated; no char or thrombus formation was observed. Compared with the control group, the study cohort had shorter mean RF application duration (26.3 ± 5.2 min vs. 89.2 ± 27.2 min; p < 0.001), shorter mean fluoroscopic time (11.2 ± 8.5 min vs. 19.5 ± 6.8 min; p < 0.001), and lower acute dormant PV reconduction (0 of 35 vs. 5 of 35; p = 0.024). At 3 months, 23 patients underwent remapping: 39 of 46 PV pairs (84.8%) remained durably isolated in 17 of these patients (73.9%). Conclusions This first-in-human series demonstrated that temperature-controlled irrigated ablation produced rapid, efficient, and durable PV isolation. (ACT DiamondTemp Temperature-Controlled and Contact Sensing RF Ablation Clinical Trial for Atrial Fibrillation [TRAC-AF]; NCT02821351)
Klasifikace
Druh
J<sub>ost</sub> - Ostatní články v recenzovaných periodicích
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of the American College of Cardiology
ISSN
0735-1097
e-ISSN
—
Svazek periodika
70
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
12
Strana od-do
542-553
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—