Retrieval of Chronically Implanted Dual-chamber Leadless Pacemakers in an Ovine Model
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F23%3A00009707" target="_blank" >RIV/00023884:_____/23:00009707 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.ahajournals.org/doi/10.1161/CIRCEP.123.012232" target="_blank" >https://www.ahajournals.org/doi/10.1161/CIRCEP.123.012232</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1161/CIRCEP.123.012232" target="_blank" >10.1161/CIRCEP.123.012232</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Retrieval of Chronically Implanted Dual-chamber Leadless Pacemakers in an Ovine Model
Popis výsledku v původním jazyce
BACKGROUND:The clinical utilization of leadless pacemakers (LPs) as an alternative to traditional transvenous pacemakers is likely to increase with the advent of dual-chamber LP systems. Since device retrieval to allow LP upgrade or replacement will become an important capability, the first such dual-chamber, helix-fixation LP system (Aveir DR; Abbott, Abbott Park, IL) was specifically designed to allow catheter-based retrieval. In this study, the preclinical performance and safety of retrieving chronically implanted dual-chamber LPs was evaluated.METHODS:Atrial and ventricular LPs were implanted in the right atrial appendage and right ventricular apex of 9 healthy ovine subjects. After approximate to 2 years, the LPs were retrieved using a dedicated transvenous retrieval catheter (Aveir Retrieval Catheter; Abbott) by snaring, docking, and unscrewing from the myocardium. Comprehensive necropsy/histopathology studies were conducted to evaluate device- and procedure-related outcomes.RESULTS:At a median of 1.9 years postimplant (range, 1.8-2.6), all 18 of 18 (100%) LPs were retrieved from 9 ovine subjects without complications. The median retrieval procedure duration for both LPs, from first-catheter-in to last-catheter-out, was 13.3 minutes (range, 2.5-36.4). Postretrieval, all right atrial, and right ventricular implant sites demonstrated minimal tissue disruption, with intact fibrous tissue limited to the distal device body. No significant device-related trauma, perforation, pericardial effusion, right heart or tricuspid valve injury, or chronic pulmonary thromboembolism were observed at necropsy.CONCLUSIONS:This preclinical study demonstrated the safe and effective retrieval of chronically implanted, helix-fixation, dual-chamber LP systems, paving the way for clinical studies of LP retrieval.
Název v anglickém jazyce
Retrieval of Chronically Implanted Dual-chamber Leadless Pacemakers in an Ovine Model
Popis výsledku anglicky
BACKGROUND:The clinical utilization of leadless pacemakers (LPs) as an alternative to traditional transvenous pacemakers is likely to increase with the advent of dual-chamber LP systems. Since device retrieval to allow LP upgrade or replacement will become an important capability, the first such dual-chamber, helix-fixation LP system (Aveir DR; Abbott, Abbott Park, IL) was specifically designed to allow catheter-based retrieval. In this study, the preclinical performance and safety of retrieving chronically implanted dual-chamber LPs was evaluated.METHODS:Atrial and ventricular LPs were implanted in the right atrial appendage and right ventricular apex of 9 healthy ovine subjects. After approximate to 2 years, the LPs were retrieved using a dedicated transvenous retrieval catheter (Aveir Retrieval Catheter; Abbott) by snaring, docking, and unscrewing from the myocardium. Comprehensive necropsy/histopathology studies were conducted to evaluate device- and procedure-related outcomes.RESULTS:At a median of 1.9 years postimplant (range, 1.8-2.6), all 18 of 18 (100%) LPs were retrieved from 9 ovine subjects without complications. The median retrieval procedure duration for both LPs, from first-catheter-in to last-catheter-out, was 13.3 minutes (range, 2.5-36.4). Postretrieval, all right atrial, and right ventricular implant sites demonstrated minimal tissue disruption, with intact fibrous tissue limited to the distal device body. No significant device-related trauma, perforation, pericardial effusion, right heart or tricuspid valve injury, or chronic pulmonary thromboembolism were observed at necropsy.CONCLUSIONS:This preclinical study demonstrated the safe and effective retrieval of chronically implanted, helix-fixation, dual-chamber LP systems, paving the way for clinical studies of LP retrieval.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2023
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Circulation-Arrhythmia and Electrophysiology
ISSN
1941-3149
e-ISSN
—
Svazek periodika
16
Číslo periodika v rámci svazku
10
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
572-578
Kód UT WoS článku
001084931400001
EID výsledku v databázi Scopus
2-s2.0-85174751358