Cardio-microcurrent device treatment for heart failure with reduced ejection fraction: Results from the C-MIC II open-label randomized controlled trial
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F25%3A00010191" target="_blank" >RIV/00023884:_____/25:00010191 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216208:11110/25:10511080
Výsledek na webu
<a href="https://onlinelibrary-wiley-com.ezproxy.lib.cas.cz/doi/full/10.1002/ejhf.3763" target="_blank" >https://onlinelibrary-wiley-com.ezproxy.lib.cas.cz/doi/full/10.1002/ejhf.3763</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/ejhf.3763" target="_blank" >10.1002/ejhf.3763</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Cardio-microcurrent device treatment for heart failure with reduced ejection fraction: Results from the C-MIC II open-label randomized controlled trial
Popis výsledku v původním jazyce
Aims In patients with heart failure, alterations in electrical fields generated within the myocardium have been associated with myocardial oedema which can act as a substrate for left ventricular dysfunction. Safety and efficacy of a direct microcurrent therapy using an implanted generator (C-MIC) remain uncertain. Methods and results Ambulatory patients with non-ischaemic dilated cardiomyopathy with left ventricular ejection fraction (LVEF) of 25% to 35% and New York Heart Association (NYHA) class III-IV were randomized to C-MIC (device) or control group in addition to guideline-directed medical therapy. The primary endpoint was change in LVEF at 6 months. Pre-specified secondary endpoints included 6-min walk distance (6MWD), Kansas City Cardiomyopathy Questionnaire overall summary score (KCCQ-OSS), and NYHA functional class. Of 70 patients randomized, 65 were included in modified intention-to-treat analysis (C-MIC device: n = 32; control: n = 33). At 6 months, treatment with C-MIC versus control improved LVEF (mean difference: 5.1%; 95% confidence interval [CI] 3.1-7.1%, p < 0.001). The proportions of patients with improvement in at least one NYHA class (risk difference: 68.9%; 95% CI 50.6-87.2, p < 0.001), an increase of >= 5 points in KCCQ-OSS (risk difference: 60.0%; 95% CI 42.3-77.6, p < 0.001), and an increase of >= 30% in 6MWD (risk difference: 38.3%; 95% CI 14.4-62.2) were substantially higher in the device versus control group (p < 0.002). Conclusions In patients with non-ischaemic chronic heart failure with reduced ejection fraction, the C-MIC device compared with control improved LVEF, symptoms, functional capacity and quality of life.
Název v anglickém jazyce
Cardio-microcurrent device treatment for heart failure with reduced ejection fraction: Results from the C-MIC II open-label randomized controlled trial
Popis výsledku anglicky
Aims In patients with heart failure, alterations in electrical fields generated within the myocardium have been associated with myocardial oedema which can act as a substrate for left ventricular dysfunction. Safety and efficacy of a direct microcurrent therapy using an implanted generator (C-MIC) remain uncertain. Methods and results Ambulatory patients with non-ischaemic dilated cardiomyopathy with left ventricular ejection fraction (LVEF) of 25% to 35% and New York Heart Association (NYHA) class III-IV were randomized to C-MIC (device) or control group in addition to guideline-directed medical therapy. The primary endpoint was change in LVEF at 6 months. Pre-specified secondary endpoints included 6-min walk distance (6MWD), Kansas City Cardiomyopathy Questionnaire overall summary score (KCCQ-OSS), and NYHA functional class. Of 70 patients randomized, 65 were included in modified intention-to-treat analysis (C-MIC device: n = 32; control: n = 33). At 6 months, treatment with C-MIC versus control improved LVEF (mean difference: 5.1%; 95% confidence interval [CI] 3.1-7.1%, p < 0.001). The proportions of patients with improvement in at least one NYHA class (risk difference: 68.9%; 95% CI 50.6-87.2, p < 0.001), an increase of >= 5 points in KCCQ-OSS (risk difference: 60.0%; 95% CI 42.3-77.6, p < 0.001), and an increase of >= 30% in 6MWD (risk difference: 38.3%; 95% CI 14.4-62.2) were substantially higher in the device versus control group (p < 0.002). Conclusions In patients with non-ischaemic chronic heart failure with reduced ejection fraction, the C-MIC device compared with control improved LVEF, symptoms, functional capacity and quality of life.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
European Journal of Heart Failure
ISSN
1388-9842
e-ISSN
—
Svazek periodika
27
Číslo periodika v rámci svazku
10
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
13
Strana od-do
1837-1849
Kód UT WoS článku
001529008900001
EID výsledku v databázi Scopus
2-s2.0-105012486940