Polymer-free sirolimus-eluting stents in a large-scale all-comers population
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00064173%3A_____%2F17%3AN0000150" target="_blank" >RIV/00064173:_____/17:N0000150 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1136/openhrt-2017-000592" target="_blank" >http://dx.doi.org/10.1136/openhrt-2017-000592</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1136/openhrt-2017-000592" target="_blank" >10.1136/openhrt-2017-000592</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Polymer-free sirolimus-eluting stents in a large-scale all-comers population
Popis výsledku v původním jazyce
Objective The objective of this study was to assess the safety and efficacy of a polymer-free sirolimus coated, ultrathin strut drug-eluting stent (PF-SES) in an unselected patient population with a focus on acute coronary syndrome (ACS). Furthermore, stable coronary artery disease (CAD) with short (≤6 months) versus long (>6 months) dual antiplatelet therapy (DAPT) were also studied. Methods Patients who received PF-SES were investigated in an unselected large-scale international, single-armed, multicenter, 'all comers' observational study. The primary endpoint was the 9-month target lesion revascularisation (TLR) rate, whereas secondary endpoints included the 9-month major adverse cardiac events (MACE) and procedural success rates. A priori defined subgroups such as patients with ACS, diabetes, lesion subsets and procedural characteristics relative to DAPT were investigated. Results A total of 2877 patients of whom 1084 had ACS were treated with PF-SES (1.31±0.75 stents per patient). At 9 months, the accumulated overall TLR rate was 2.3% (58/2513). There was no significant difference between ACS and stable CAD (2.6% vs 2.1%, p=0.389). However, the overall MACE rate was 4.3% (108/2513) with a higher rate in patients with ACS when compared with the stable CAD subgroup (6.1%, 58/947 vs 3.2%, 50/1566, p<0.001). Conclusions PF-SES angioplasty is safe and effective in the daily clinical routine with low rates of TLR and MACE in an unselected patient population. Our data are in agreement with prior clinical findings that extended DAPT duration beyond 6 months do not improve clinical outcomes in patients with stable CAD (ClinicalTrials.gov Identifier NCT02629575). Trial registration number NCT02629575.
Název v anglickém jazyce
Polymer-free sirolimus-eluting stents in a large-scale all-comers population
Popis výsledku anglicky
Objective The objective of this study was to assess the safety and efficacy of a polymer-free sirolimus coated, ultrathin strut drug-eluting stent (PF-SES) in an unselected patient population with a focus on acute coronary syndrome (ACS). Furthermore, stable coronary artery disease (CAD) with short (≤6 months) versus long (>6 months) dual antiplatelet therapy (DAPT) were also studied. Methods Patients who received PF-SES were investigated in an unselected large-scale international, single-armed, multicenter, 'all comers' observational study. The primary endpoint was the 9-month target lesion revascularisation (TLR) rate, whereas secondary endpoints included the 9-month major adverse cardiac events (MACE) and procedural success rates. A priori defined subgroups such as patients with ACS, diabetes, lesion subsets and procedural characteristics relative to DAPT were investigated. Results A total of 2877 patients of whom 1084 had ACS were treated with PF-SES (1.31±0.75 stents per patient). At 9 months, the accumulated overall TLR rate was 2.3% (58/2513). There was no significant difference between ACS and stable CAD (2.6% vs 2.1%, p=0.389). However, the overall MACE rate was 4.3% (108/2513) with a higher rate in patients with ACS when compared with the stable CAD subgroup (6.1%, 58/947 vs 3.2%, 50/1566, p<0.001). Conclusions PF-SES angioplasty is safe and effective in the daily clinical routine with low rates of TLR and MACE in an unselected patient population. Our data are in agreement with prior clinical findings that extended DAPT duration beyond 6 months do not improve clinical outcomes in patients with stable CAD (ClinicalTrials.gov Identifier NCT02629575). Trial registration number NCT02629575.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Open Heart
ISSN
2053-3624
e-ISSN
2053-3624
Svazek periodika
4
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
10
Strana od-do
e000592
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85020315571