Unexpected ischemia during transcatheter patent arterial duct closure
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00064203%3A_____%2F19%3A10400170" target="_blank" >RIV/00064203:_____/19:10400170 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216208:11130/19:10400170
Výsledek na webu
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=tawMjQI_Kp" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=tawMjQI_Kp</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1017/S104795111900218X" target="_blank" >10.1017/S104795111900218X</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Unexpected ischemia during transcatheter patent arterial duct closure
Popis výsledku v původním jazyce
A child undergoing routine transcatheter patent arterial duct closure developed severe transient ischemic changes in the electrocardiogram (Pardee waves) while the aortic retention skirt of the AmplatzerTM Duct Occluder was pulled against the duct orifice. The occluder was then released, and the delivery system was pulled back to inferior caval vein which led to electrocardiogram normalisation. Aortic root angiography showed a single coronary artery originating from the right sinus of Valsalva with the left coronary stem wedged behind the posterior aspect of the right ventricular outflow tract. We believe that the left coronary artery was compressed while applying tension on the occluder delivery system. (C) Cambridge University Press 2019.
Název v anglickém jazyce
Unexpected ischemia during transcatheter patent arterial duct closure
Popis výsledku anglicky
A child undergoing routine transcatheter patent arterial duct closure developed severe transient ischemic changes in the electrocardiogram (Pardee waves) while the aortic retention skirt of the AmplatzerTM Duct Occluder was pulled against the duct orifice. The occluder was then released, and the delivery system was pulled back to inferior caval vein which led to electrocardiogram normalisation. Aortic root angiography showed a single coronary artery originating from the right sinus of Valsalva with the left coronary stem wedged behind the posterior aspect of the right ventricular outflow tract. We believe that the left coronary artery was compressed while applying tension on the occluder delivery system. (C) Cambridge University Press 2019.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Cardiology in the Young
ISSN
1047-9511
e-ISSN
—
Svazek periodika
29
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
2
Strana od-do
1410-1411
Kód UT WoS článku
000511340100019
EID výsledku v databázi Scopus
2-s2.0-85072670869