Magnetizable Duplex Steel Stents Enable Endothelial Cell Capture
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00159816%3A_____%2F13%3A00060589" target="_blank" >RIV/00159816:_____/13:00060589 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/TMAG.2012.2224326" target="_blank" >http://dx.doi.org/10.1109/TMAG.2012.2224326</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/TMAG.2012.2224326" target="_blank" >10.1109/TMAG.2012.2224326</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Magnetizable Duplex Steel Stents Enable Endothelial Cell Capture
Popis výsledku v původním jazyce
Emerging medical nanotechnology applications often utilize magnetic forces to guide the movement of superparamagnetic particle linked cells and drugs in order to achieve a therapeutic effect. Superparamagnetic particle labeled endothelial cells have previously been captured on the surface of prototype nickel-plated stents in proof of concept studies. Facilitated endothelialization may help improve the healing of stented arteries and reduce the risk of stent thrombosis and restenosis. Extensive evaluation of candidate materials led to the development of a magnetizable 2205 duplex stainless steel stent. Magnetic field strengths of approximately 630 mG were induced within these stents by holding them in close proximity to a 0.7 T rare earth magnet. The magnetic field strength was reliably maintained over several days, but was partially reduced upon mild mechanical shock or plastic deformation. Mechanical testing demonstrated that stents could withstand crimping and expansion necessary for
Název v anglickém jazyce
Magnetizable Duplex Steel Stents Enable Endothelial Cell Capture
Popis výsledku anglicky
Emerging medical nanotechnology applications often utilize magnetic forces to guide the movement of superparamagnetic particle linked cells and drugs in order to achieve a therapeutic effect. Superparamagnetic particle labeled endothelial cells have previously been captured on the surface of prototype nickel-plated stents in proof of concept studies. Facilitated endothelialization may help improve the healing of stented arteries and reduce the risk of stent thrombosis and restenosis. Extensive evaluation of candidate materials led to the development of a magnetizable 2205 duplex stainless steel stent. Magnetic field strengths of approximately 630 mG were induced within these stents by holding them in close proximity to a 0.7 T rare earth magnet. The magnetic field strength was reliably maintained over several days, but was partially reduced upon mild mechanical shock or plastic deformation. Mechanical testing demonstrated that stents could withstand crimping and expansion necessary for
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BO - Biofyzika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/ED1.100%2F02%2F0123" target="_blank" >ED1.100/02/0123: Fakultní nemocnice u sv. Anny v Brně - Mezinárodní centrum klinického výzkumu (FNUSA - ICRC)</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Magnetics
ISSN
0018-9464
e-ISSN
—
Svazek periodika
49
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
4
Strana od-do
463-466
Kód UT WoS článku
000312831500052
EID výsledku v databázi Scopus
—