Results from the Tack Optimized Balloon Angioplasty (TOBA) study demonstrate the benefits of minimal metal implants for dissection repair after angioplasty
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00159816%3A_____%2F16%3A00065354" target="_blank" >RIV/00159816:_____/16:00065354 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216224:14110/16:00090354
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.jvs.2016.02.043" target="_blank" >http://dx.doi.org/10.1016/j.jvs.2016.02.043</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.jvs.2016.02.043" target="_blank" >10.1016/j.jvs.2016.02.043</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Results from the Tack Optimized Balloon Angioplasty (TOBA) study demonstrate the benefits of minimal metal implants for dissection repair after angioplasty
Popis výsledku v původním jazyce
Objective: The mechanism of angioplasty is disruption of atherosclerotic plaque, which often results in dissections. Dissection after percutaneous transluminal angioplasty (PTA) remains a significant clinical problem and untreated may cause acute occlusion or later restenosis. Stents are used to manage dissections, which is often followed by in-stent restenosis and occasionally stent fracture. Tack (Intact Vascular, Wayne, Pa) implants have minimal metal and low radial force and are specifically designed for dissection repair. This study evaluated Tack implants for treatment of dissections resulting from standard balloon PTA for femoral-popliteal arterial disease. Twelve-month outcomes after Tack treatment of post-PTA dissections are described. Methods: This prospective, single-arm study evaluated patients with ischemia (Rutherford clinical category 2-4) caused by lesions of the superficial femoral and popliteal arteries. Patients were treated with standard balloon angioplasty, and post-PTA dissections were treated with Tacks. The primary end points were core laboratory-adjudicated device technical success, defined as the ability of the Tack implants to resolve post-PTA dissection, and device safety, defined as the absence of new-onset major adverse events. Patients were followed up to 12 months after implantation. Results: Tacks were used in 130 patients with post-PTA dissections (74.0% }= grade C). Technical success was achieved in 128 (98.5%) patients with no major adverse events at 30 days. The 12-month patency was 76.4%, and freedom from target lesion revascularization was 89.5%. Significant improvement from baseline was observed in Rutherford clinical category (82.8% with grade {= 1) and ankle-brachial index (0.68 +/- 0.18 to 0.94 +/- 0.15; P<.0001). Conclusions: Tack implant treatment of post-PTA dissection was safe, produced reasonable patency, and resulted in low rates of target lesion revascularization. (...)
Název v anglickém jazyce
Results from the Tack Optimized Balloon Angioplasty (TOBA) study demonstrate the benefits of minimal metal implants for dissection repair after angioplasty
Popis výsledku anglicky
Objective: The mechanism of angioplasty is disruption of atherosclerotic plaque, which often results in dissections. Dissection after percutaneous transluminal angioplasty (PTA) remains a significant clinical problem and untreated may cause acute occlusion or later restenosis. Stents are used to manage dissections, which is often followed by in-stent restenosis and occasionally stent fracture. Tack (Intact Vascular, Wayne, Pa) implants have minimal metal and low radial force and are specifically designed for dissection repair. This study evaluated Tack implants for treatment of dissections resulting from standard balloon PTA for femoral-popliteal arterial disease. Twelve-month outcomes after Tack treatment of post-PTA dissections are described. Methods: This prospective, single-arm study evaluated patients with ischemia (Rutherford clinical category 2-4) caused by lesions of the superficial femoral and popliteal arteries. Patients were treated with standard balloon angioplasty, and post-PTA dissections were treated with Tacks. The primary end points were core laboratory-adjudicated device technical success, defined as the ability of the Tack implants to resolve post-PTA dissection, and device safety, defined as the absence of new-onset major adverse events. Patients were followed up to 12 months after implantation. Results: Tacks were used in 130 patients with post-PTA dissections (74.0% }= grade C). Technical success was achieved in 128 (98.5%) patients with no major adverse events at 30 days. The 12-month patency was 76.4%, and freedom from target lesion revascularization was 89.5%. Significant improvement from baseline was observed in Rutherford clinical category (82.8% with grade {= 1) and ankle-brachial index (0.68 +/- 0.18 to 0.94 +/- 0.15; P<.0001). Conclusions: Tack implant treatment of post-PTA dissection was safe, produced reasonable patency, and resulted in low rates of target lesion revascularization. (...)
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FJ - Chirurgie včetně transplantologie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Vascular Surgery
ISSN
0741-5214
e-ISSN
—
Svazek periodika
64
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
109-116
Kód UT WoS článku
000378562900016
EID výsledku v databázi Scopus
—