Focal control of non-invasive deep brain stimulation using multipolar temporal interference
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00159816%3A_____%2F25%3A00082300" target="_blank" >RIV/00159816:_____/25:00082300 - isvavai.cz</a>
Výsledek na webu
<a href="https://link.springer.com/article/10.1186/s42234-025-00169-6" target="_blank" >https://link.springer.com/article/10.1186/s42234-025-00169-6</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1186/s42234-025-00169-6" target="_blank" >10.1186/s42234-025-00169-6</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Focal control of non-invasive deep brain stimulation using multipolar temporal interference
Popis výsledku v původním jazyce
Temporal interference (TI) is a method of non-invasive brain stimulation using transcutaneous electrodes which allows the targeting and modulation of deeper brain structures, not normally associated with non-invasive simulation, while avoiding unwanted stimulation of shallower cortical structures. The properties of TI have been previously demonstrated, however, the problem of decoupling stimulation focality from stimulation intensity has not yet been well addressed. In this paper, we provide a possible novel solution, multipolar TI (mTI), which allows increased independent control over both the size of the stimulated region and the stimulation intensity. The mTI method uses multiple carrier frequencies to create multiple overlapping amplitude-modulated envelopes, rather than using one envelope as in standard TI. The study presents an explanation of the concept of mTI along with experimental data gathered from Rhesus macaques and mice. We improved the focality at depth in anesthetized mice and monkeys, and using the new focality in awake monkeys, evoked targeted activity at depth in the superior colliculus. The mTI method could be an interesting and potentially useful new tool alongside other forms of non-invasive brain stimulation. Teaser Multipolar Temporal Interference Stimulation can produce a more focal brain stimulation at depth compared to Temporal Interference.
Název v anglickém jazyce
Focal control of non-invasive deep brain stimulation using multipolar temporal interference
Popis výsledku anglicky
Temporal interference (TI) is a method of non-invasive brain stimulation using transcutaneous electrodes which allows the targeting and modulation of deeper brain structures, not normally associated with non-invasive simulation, while avoiding unwanted stimulation of shallower cortical structures. The properties of TI have been previously demonstrated, however, the problem of decoupling stimulation focality from stimulation intensity has not yet been well addressed. In this paper, we provide a possible novel solution, multipolar TI (mTI), which allows increased independent control over both the size of the stimulated region and the stimulation intensity. The mTI method uses multiple carrier frequencies to create multiple overlapping amplitude-modulated envelopes, rather than using one envelope as in standard TI. The study presents an explanation of the concept of mTI along with experimental data gathered from Rhesus macaques and mice. We improved the focality at depth in anesthetized mice and monkeys, and using the new focality in awake monkeys, evoked targeted activity at depth in the superior colliculus. The mTI method could be an interesting and potentially useful new tool alongside other forms of non-invasive brain stimulation. Teaser Multipolar Temporal Interference Stimulation can produce a more focal brain stimulation at depth compared to Temporal Interference.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
30103 - Neurosciences (including psychophysiology)
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Bioelectronic medicine
ISSN
—
e-ISSN
2332-8886
Svazek periodika
11
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
15
Strana od-do
nestránkováno
Kód UT WoS článku
001484107600001
EID výsledku v databázi Scopus
2-s2.0-105001282940