International comparisons of He-Ne lasers stabilized with 127I2 at ?= 633 nm. Comparison of the use of the fifth and third harmonic locking techniques
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F00%3A%230000131" target="_blank" >RIV/00177016:_____/00:#0000131 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
International comparisons of He-Ne lasers stabilized with 127I2 at ?= 633 nm. Comparison of the use of the fifth and third harmonic locking techniques
Popis výsledku v původním jazyce
An international comparison of 127I2-stabilized 633 nm He-Ne lasers locked either by the fifth- or the third-harmonic technique was carried out at the Finnish Centre for Metrology and Accreditation (MIKES) in September 1998. The participating laboratories were the MIKES; the Metrology Research Institute, Helsinki University of Technology (HUT); the Chinese National Institute of Metrology (NIM); the Czech Metrological Institute (CMI); the D. I. Mendeleyev Institute for Metrology (VNIIM); and the Bureau International des Poids et Mesures (BIPM). Raw measured frequency differences between lasers were in the following ranges: -5.2 kHz to +11.9 kHz and -5.8 kHz to +4.9 kHz for the third- and fifth-harmonic locking techniques, respectively. The frequency offsets between the third- and fifth-harmonic techniques for different components (d to g) were in the range 26 kHz to 35 kHz.
Název v anglickém jazyce
International comparisons of He-Ne lasers stabilized with 127I2 at ?= 633 nm. Comparison of the use of the fifth and third harmonic locking techniques
Popis výsledku anglicky
An international comparison of 127I2-stabilized 633 nm He-Ne lasers locked either by the fifth- or the third-harmonic technique was carried out at the Finnish Centre for Metrology and Accreditation (MIKES) in September 1998. The participating laboratories were the MIKES; the Metrology Research Institute, Helsinki University of Technology (HUT); the Chinese National Institute of Metrology (NIM); the Czech Metrological Institute (CMI); the D. I. Mendeleyev Institute for Metrology (VNIIM); and the Bureau International des Poids et Mesures (BIPM). Raw measured frequency differences between lasers were in the following ranges: -5.2 kHz to +11.9 kHz and -5.8 kHz to +4.9 kHz for the third- and fifth-harmonic locking techniques, respectively. The frequency offsets between the third- and fifth-harmonic techniques for different components (d to g) were in the range 26 kHz to 35 kHz.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
R - Projekt Ramcoveho programu EK
Ostatní
Rok uplatnění
2000
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Metrologia
ISSN
1681-7575
e-ISSN
—
Svazek periodika
37
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—