Predictable quantum efficient detector: II. Characterization and confirmed responsivity
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F13%3A%230000959" target="_blank" >RIV/00177016:_____/13:#0000959 - isvavai.cz</a>
Výsledek na webu
<a href="http://iopscience.iop.org/0026-1394/50/4/395/article" target="_blank" >http://iopscience.iop.org/0026-1394/50/4/395/article</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/0026-1394/50/4/395" target="_blank" >10.1088/0026-1394/50/4/395</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Predictable quantum efficient detector: II. Characterization and confirmed responsivity
Popis výsledku v původním jazyce
The predictable quantum efficient detector (PQED) is intended to become a new primary standard for radiant power measurements in the wavelength range from 400 nm to 800 nm. Characterization results of custom-made single induced junction photodiodes as they are used in the PQED and of assembled PQEDs are presented. The single photodiodes were tested in terms of linearity and spatial uniformity of the spectral responsivity. The highly uniform photodiodes were proved to be linear over seven orders of magnitude, i.e. in the radiant power range from 100 pW to 400 ?W. The assembled PQED has been compared with a cryogenic electrical substitution radiometer with a very low uncertainty of the order of 30 ppm. Experimental results show good agreement with the modelled response of the PQED to optical radiation and prove a near unity external quantum efficiency.
Název v anglickém jazyce
Predictable quantum efficient detector: II. Characterization and confirmed responsivity
Popis výsledku anglicky
The predictable quantum efficient detector (PQED) is intended to become a new primary standard for radiant power measurements in the wavelength range from 400 nm to 800 nm. Characterization results of custom-made single induced junction photodiodes as they are used in the PQED and of assembled PQEDs are presented. The single photodiodes were tested in terms of linearity and spatial uniformity of the spectral responsivity. The highly uniform photodiodes were proved to be linear over seven orders of magnitude, i.e. in the radiant power range from 100 pW to 400 ?W. The assembled PQED has been compared with a cryogenic electrical substitution radiometer with a very low uncertainty of the order of 30 ppm. Experimental results show good agreement with the modelled response of the PQED to optical radiation and prove a near unity external quantum efficiency.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/7AX13029" target="_blank" >7AX13029: New primary standards and traceability for radiometry</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Metrologia
ISSN
1681-7575
e-ISSN
—
Svazek periodika
50
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
FR - Francouzská republika
Počet stran výsledku
7
Strana od-do
395-401
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—