Modeling of thermal profile inside guarded hot plate apparatus
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F13%3A%230001009" target="_blank" >RIV/00177016:_____/13:#0001009 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Modeling of thermal profile inside guarded hot plate apparatus
Popis výsledku v původním jazyce
Guarded hot plate (GHP) method is widely accepted for determination of heat-transfer properties of thermo-insulating materials. There is an effort to extend GHP measurements to a high temperature region keeping acceptable accuracy. High temperatures represent higher demands for materials used. Also great attention is paid to own apparatus design to avoid undesirable heat losses. The guarded hot plate apparatus is designed to provide unidirectional heat flow from hot plate to cold plate through a specimen enabling the simple determination of desired heat transfer quantity (thermal conductivity, thermal resistance etc.). The goals of modeling temperature profile inside GHP apparatus are to confirm unidirectional character of heat flow in the metering section through the specimen and follow time evolution of the temperature profile estimating time needed to reach a steady state.
Název v anglickém jazyce
Modeling of thermal profile inside guarded hot plate apparatus
Popis výsledku anglicky
Guarded hot plate (GHP) method is widely accepted for determination of heat-transfer properties of thermo-insulating materials. There is an effort to extend GHP measurements to a high temperature region keeping acceptable accuracy. High temperatures represent higher demands for materials used. Also great attention is paid to own apparatus design to avoid undesirable heat losses. The guarded hot plate apparatus is designed to provide unidirectional heat flow from hot plate to cold plate through a specimen enabling the simple determination of desired heat transfer quantity (thermal conductivity, thermal resistance etc.). The goals of modeling temperature profile inside GHP apparatus are to confirm unidirectional character of heat flow in the metering section through the specimen and follow time evolution of the temperature profile estimating time needed to reach a steady state.
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
BJ - Termodynamika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/7AX13026" target="_blank" >7AX13026: Metrology for thermal protection materials</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů