CALIBRATION OF A MEASURING DEVICE T O DETERMINE SPECIFIC SATURATION MAGNETIZATION BY A SPECIAL COIL
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F14%3A%230000831" target="_blank" >RIV/00177016:_____/14:#0000831 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.2478/jee-2014-0041" target="_blank" >http://dx.doi.org/10.2478/jee-2014-0041</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.2478/jee-2014-0041" target="_blank" >10.2478/jee-2014-0041</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
CALIBRATION OF A MEASURING DEVICE T O DETERMINE SPECIFIC SATURATION MAGNETIZATION BY A SPECIAL COIL
Popis výsledku v původním jazyce
Specific saturation magnetization ?S is an important parameter of ferromagnetic materials. It is in principle independent of the structure and shape of the sample. There are two ways to calibrate devices for determining ?S: the standard method, using a calibrated magnetic moment standard, or an alternative method, using a special moment coil. This paper presents a calibration method used at Czech Metrology Institute for calibration of Koerzimat 1.096 using a special moment coil. Special attention is given to a description of the special moment coils used for Koerzimat 1.096 calibration, and to analysis of sources of uncertainty. Calibration expanded uncertainty of 0.4 up to 0.6 % can be achieved using this method.
Název v anglickém jazyce
CALIBRATION OF A MEASURING DEVICE T O DETERMINE SPECIFIC SATURATION MAGNETIZATION BY A SPECIAL COIL
Popis výsledku anglicky
Specific saturation magnetization ?S is an important parameter of ferromagnetic materials. It is in principle independent of the structure and shape of the sample. There are two ways to calibrate devices for determining ?S: the standard method, using a calibrated magnetic moment standard, or an alternative method, using a special moment coil. This paper presents a calibration method used at Czech Metrology Institute for calibration of Koerzimat 1.096 using a special moment coil. Special attention is given to a description of the special moment coils used for Koerzimat 1.096 calibration, and to analysis of sources of uncertainty. Calibration expanded uncertainty of 0.4 up to 0.6 % can be achieved using this method.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of E LECTRICAL ENGINEERING,
ISSN
1335-3632
e-ISSN
—
Svazek periodika
65
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
SK - Slovenská republika
Počet stran výsledku
3
Strana od-do
262-264
Kód UT WoS článku
000342874700011