Calibration of Industrial Platinum Resistance Thermometers up to 700 °C
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F14%3A%230001020" target="_blank" >RIV/00177016:_____/14:#0001020 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Calibration of Industrial Platinum Resistance Thermometers up to 700 °C
Popis výsledku v původním jazyce
Industrial grade platinum resistance thermometers were calibrated in the temperature range from 200°C to 700°C . Both wire-wound and thin-film sensor-based thermometers were investigated. The purpose of the study was to investigate thermometers which could be used in future coal power plants. The calibrations were performed in a vertical cesium heat-pipe furnace and in a horizontal and vertical sodium heat-pipe furnace. The reference thermometer was a standard platinum resistance thermometer calibratedat fixed points up to the aluminum point. In addition to calibration, various thermal tests including immersion measurements and thermal-cycling tests were performed. The stability of the sensors was determined by monitoring the ice-point resistance. Possible contamination of the sensors was determined by measuring the resistance ratio R(30°C)/R(10°C) several times during the measurement period. The calibration curves were compared with the ICE 60751 standard and International Temperatur
Název v anglickém jazyce
Calibration of Industrial Platinum Resistance Thermometers up to 700 °C
Popis výsledku anglicky
Industrial grade platinum resistance thermometers were calibrated in the temperature range from 200°C to 700°C . Both wire-wound and thin-film sensor-based thermometers were investigated. The purpose of the study was to investigate thermometers which could be used in future coal power plants. The calibrations were performed in a vertical cesium heat-pipe furnace and in a horizontal and vertical sodium heat-pipe furnace. The reference thermometer was a standard platinum resistance thermometer calibratedat fixed points up to the aluminum point. In addition to calibration, various thermal tests including immersion measurements and thermal-cycling tests were performed. The stability of the sensors was determined by monitoring the ice-point resistance. Possible contamination of the sensors was determined by measuring the resistance ratio R(30°C)/R(10°C) several times during the measurement period. The calibration curves were compared with the ICE 60751 standard and International Temperatur
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BJ - Termodynamika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/7A11082" target="_blank" >7A11082: Metrology for improved power plant efficiency</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
International Jurnal of Thermophysics
ISSN
1572-9567
e-ISSN
—
Svazek periodika
35
Číslo periodika v rámci svazku
3-4
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
12
Strana od-do
668-680