Setup for generating an AC magnetic field from 3 kHz up to 100 kHz
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F15%3A%230000832" target="_blank" >RIV/00177016:_____/15:#0000832 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/TMAG.2014.2359296" target="_blank" >http://dx.doi.org/10.1109/TMAG.2014.2359296</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/TMAG.2014.2359296" target="_blank" >10.1109/TMAG.2014.2359296</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Setup for generating an AC magnetic field from 3 kHz up to 100 kHz
Popis výsledku v původním jazyce
Generating a stable, homogeneous reference ac magnetic field is important for traceable calibrations of ac magnetic field analyzers with a triaxial coil probe (e.g., EFA 300, C.A. 42). These probes are widely used by health and safety professionals, in manufacturing, and in service industries. This paper presents a single-layer Helmholtz-type solenoid for generating the reference ac magnetic field in the frequency range from 3 to 100 kHz. Special attention is given to a full characterization of this setup, including a comparison between the analytical and finite element method design of the solenoid, on the one hand, and experimental results, on the other. The magnetic flux density inside the solenoid can be generated with expanded uncertainty of (0.4%?0.8 %) for k = 2 using a special search coil.
Název v anglickém jazyce
Setup for generating an AC magnetic field from 3 kHz up to 100 kHz
Popis výsledku anglicky
Generating a stable, homogeneous reference ac magnetic field is important for traceable calibrations of ac magnetic field analyzers with a triaxial coil probe (e.g., EFA 300, C.A. 42). These probes are widely used by health and safety professionals, in manufacturing, and in service industries. This paper presents a single-layer Helmholtz-type solenoid for generating the reference ac magnetic field in the frequency range from 3 to 100 kHz. Special attention is given to a full characterization of this setup, including a comparison between the analytical and finite element method design of the solenoid, on the one hand, and experimental results, on the other. The magnetic flux density inside the solenoid can be generated with expanded uncertainty of (0.4%?0.8 %) for k = 2 using a special search coil.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE TRANSACTIONS ON MAGNETICS
ISSN
0018-9464
e-ISSN
—
Svazek periodika
51
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
4
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—