Calibration of freeform standard
Popis výsledku
Identifikátory výsledku
Kód výsledku v IS VaVaI
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Calibration of freeform standard
Popis výsledku v původním jazyce
A new freeform reference standard has been developed and manufactured in cooperation of CMI (Czech Metrology Institute) and CTU in Prague (Czech Technical University) for testing freeform measurement capability of coordinate measuring machines. Geometrical-mathematical approach to design the standard has been applied and a surface of hyperbolic paraboloid (basic feature of the standard) has been chosen as a challenging surface for freeform measurement. The surface of hyperbolic paraboloid can be considered not only as a freeform surface but also as a set of 3D straight lines. Specially, the mathematical model used to design the standard has been developed in such a way that the surface of hyperbolic paraboloid contains two sets of straight lines. Two different calibration strategies have been prepared and realised by tactile measurement on coordinate measuring machine SIP CMM 5. According to the first calibration strategy corresponding to the freeform measurement commonly used, surface
Název v anglickém jazyce
Calibration of freeform standard
Popis výsledku anglicky
A new freeform reference standard has been developed and manufactured in cooperation of CMI (Czech Metrology Institute) and CTU in Prague (Czech Technical University) for testing freeform measurement capability of coordinate measuring machines. Geometrical-mathematical approach to design the standard has been applied and a surface of hyperbolic paraboloid (basic feature of the standard) has been chosen as a challenging surface for freeform measurement. The surface of hyperbolic paraboloid can be considered not only as a freeform surface but also as a set of 3D straight lines. Specially, the mathematical model used to design the standard has been developed in such a way that the surface of hyperbolic paraboloid contains two sets of straight lines. Two different calibration strategies have been prepared and realised by tactile measurement on coordinate measuring machine SIP CMM 5. According to the first calibration strategy corresponding to the freeform measurement commonly used, surface
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BA - Obecná matematika
OECD FORD obor
—
Návaznosti výsledku
Projekt
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 15th international conference of the european society for precision engineering and nanotechnology
ISBN
978-0-9566790-7-9
ISSN
—
e-ISSN
—
Počet stran výsledku
2
Strana od-do
147-148
Název nakladatele
Sieca Repro
Místo vydání
The Netherlands
Místo konání akce
Leuven, Belgium
Datum konání akce
1. 6. 2015
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—
Základní informace
Druh výsledku
D - Stať ve sborníku
CEP
BA - Obecná matematika
Rok uplatnění
2015