Characterization of the Graphite Moderated Thermal Neutron Field at CMI
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F17%3AN0000054" target="_blank" >RIV/00177016:_____/17:N0000054 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1093/rpd/ncx190" target="_blank" >https://doi.org/10.1093/rpd/ncx190</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncx190" target="_blank" >10.1093/rpd/ncx190</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Characterization of the Graphite Moderated Thermal Neutron Field at CMI
Popis výsledku v původním jazyce
The graphite pile was set up at CMI to provide reference thermal neutron field for metrology and dosimetry purposes. It consists of three Pu–Be and three Am–Be radionuclide sources located in a 1.95 m (width) × 1.95 m (length) × 2.0m (height) graphite moderator block. The neutron field in the volume of the experimental channel with dimensions of 40 cm × 40 cm × 135 cm (depth) was characterised experimentally and by Monte-Carlo calculations. Neutron fluence was determined by the activation of gold foils and Manganese tablets irradiated in nine different positions and by two 3He detectors of different construction irradiated in one position (bare and covered in three spherical PE moderators of 3, 3.5 and 4 inch diameter). Weighted mean value of the neutron fluence rate measured by all the abovementioned detectors in the pile centre was used as a calibration factor for the Monte-Carlo model predictions. The neutron spectral fluence rate calculated by validated Monte-Carlo model was used to determine the conventionally true ambient and personal dose equivalent rates at different positions.
Název v anglickém jazyce
Characterization of the Graphite Moderated Thermal Neutron Field at CMI
Popis výsledku anglicky
The graphite pile was set up at CMI to provide reference thermal neutron field for metrology and dosimetry purposes. It consists of three Pu–Be and three Am–Be radionuclide sources located in a 1.95 m (width) × 1.95 m (length) × 2.0m (height) graphite moderator block. The neutron field in the volume of the experimental channel with dimensions of 40 cm × 40 cm × 135 cm (depth) was characterised experimentally and by Monte-Carlo calculations. Neutron fluence was determined by the activation of gold foils and Manganese tablets irradiated in nine different positions and by two 3He detectors of different construction irradiated in one position (bare and covered in three spherical PE moderators of 3, 3.5 and 4 inch diameter). Weighted mean value of the neutron fluence rate measured by all the abovementioned detectors in the pile centre was used as a calibration factor for the Monte-Carlo model predictions. The neutron spectral fluence rate calculated by validated Monte-Carlo model was used to determine the conventionally true ambient and personal dose equivalent rates at different positions.
Klasifikace
Druh
J<sub>ost</sub> - Ostatní články v recenzovaných periodicích
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
1742-3406
Svazek periodika
"neuveden"
Číslo periodika v rámci svazku
September 2017
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
3
Strana od-do
"nestrankovano"
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—