Nano-scale traceable magnetic field measurements - Final publishable report
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F19%3AN0000175" target="_blank" >RIV/00177016:_____/19:N0000175 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.euramet.org/research-innovation/search-research-projects/details/project/nano-scale-traceable-magnetic-field-measurements" target="_blank" >https://www.euramet.org/research-innovation/search-research-projects/details/project/nano-scale-traceable-magnetic-field-measurements</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Nano-scale traceable magnetic field measurements - Final publishable report
Popis výsledku v původním jazyce
The overall goal of this project was to develop and provide coordinated and sustainable European metrology capabilities that extend reliable and traceable measurements of spatially resolved magnetic fields down to the micrometre and nanometre length scale. The project delivered traceable scanning magnetic field microscopy and magneto optical indicator film microscopy tools with spatial resolution from 50 μm down to 500 nm. Furthermore, the project delivered validated calibration technique for magnetic force microscopy calibration with spatial resolution below 50 nm together with calibration artefacts suitable for on-site calibrations. Additionally, the project heavily contributed to the development of the first international standard for traceable nano-scale magnetic field measurements.
Název v anglickém jazyce
Nano-scale traceable magnetic field measurements - Final publishable report
Popis výsledku anglicky
The overall goal of this project was to develop and provide coordinated and sustainable European metrology capabilities that extend reliable and traceable measurements of spatially resolved magnetic fields down to the micrometre and nanometre length scale. The project delivered traceable scanning magnetic field microscopy and magneto optical indicator film microscopy tools with spatial resolution from 50 μm down to 500 nm. Furthermore, the project delivered validated calibration technique for magnetic force microscopy calibration with spatial resolution below 50 nm together with calibration artefacts suitable for on-site calibrations. Additionally, the project heavily contributed to the development of the first international standard for traceable nano-scale magnetic field measurements.
Klasifikace
Druh
V<sub>souhrn</sub> - Souhrnná výzkumná zpráva
CEP obor
—
OECD FORD obor
21002 - Nano-processes (applications on nano-scale); (biomaterials to be 2.9)
Návaznosti výsledku
Projekt
<a href="/cs/project/8B16012" target="_blank" >8B16012: Nano-scale traceable magnetic field measurements</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Počet stran výsledku
28
Místo vydání
—
Název nakladatele resp. objednatele
—
Verze
—